วิธีการทดLab Procedure - Sectioning
1. Place the sample on the slotted Table so that it is at the center of the swing of the head.
2. Place the Locking Lever in the forward position to allow the Control Rod to be moved to the furthest position.
Move the Rear Vise section to contact the rear face of the sample. Move the Front Section to a position near
the front face of the sample.
3. Tighten the Nuts that secure the Vise sections to the Table.
4. Push the Control Rod in to make firm contact with the sample and pull back the Locking Level to secure the
sample.
5. Turn on the coolant; it should flow readily to both cool and clean the cutting area.
6. Apply steady moderate pressure until the specimen is sectioned; avoid jerking motions that may cause the
blade to splinter.ลอง