This is the first phase of each test, and it is automatically performed before any subsequent manoeuvre (sRaw, TGV or
SVC/IC)
1. Wait until the end of the thermal equilibrium phase (the duration of this phase is defined in the Body Box settings). In
this first phase, the operator is waiting for the thermal equilibrium between the inside and the outside of the Body
Box. In the meanwhile the patient will begin to relax inside the box, and he/she can breathe within the PNT, in order
to familiarize themselves with the device and the technique.
During this phase, all the buttons are disabled, and the vent valve between the inside and the outside of the Body Box
is open, to allow reaching the equilibrium.
In this phase, the Body Box pressure column oscillates around the zero. Under the volume/time graph a countdown is
shown.
This is the first phase of each test, and it is automatically performed before any subsequent manoeuvre (sRaw, TGV orSVC/IC)1. Wait until the end of the thermal equilibrium phase (the duration of this phase is defined in the Body Box settings). Inthis first phase, the operator is waiting for the thermal equilibrium between the inside and the outside of the BodyBox. In the meanwhile the patient will begin to relax inside the box, and he/she can breathe within the PNT, in orderto familiarize themselves with the device and the technique.During this phase, all the buttons are disabled, and the vent valve between the inside and the outside of the Body Boxis open, to allow reaching the equilibrium.In this phase, the Body Box pressure column oscillates around the zero. Under the volume/time graph a countdown isshown.