The [Disp] key selects the manner that the value of a component will be displayed. If the Sorting mode is disabled, pressing [Disp] cycles through the following display types:
á The value being measured
á The deviation of the value from the current value – The symbol next to the measured parameter indicates that this function is active.
á The percent of deviation from the current value - The symbol next to the measured parameter indicates that this function is active.
In Sorting Mode pressing [Disp] cycles through the following display types:
á The value being measured
á The percent of deviation from a stored value - The symbol next to the measured parameter indicates that this function is active. In this mode, a PASS/FAIL message is shown in the second line of the display according to the measured deviation and the selected tolerance.
Zero/Range
The [Zero/Range] key allows access to open/short compensation. Pressing the [Zero/Range] key displays the zeroing options on the second line of the LCD – Open Short Exit. This option will zero the LCR meter for the currently selected test frequency. Pressing the [Zero/Range] for more than two seconds displays the zeroing options – OPEN SHORT Exit (note that the options are in all capital letters). This option performs an open/short calibration through the entire range of test frequencies.