separation device was a HP 6890 chromatograph (Agilent Technologies, S การแปล - separation device was a HP 6890 chromatograph (Agilent Technologies, S ไทย วิธีการพูด

separation device was a HP 6890 chr

separation device was a HP 6890 chromatograph (Agilent Technologies, Santa Clara, USA) equipped with a Flame Ionisation Detector (FID) system kept at 250 °C and a polar DB-Wax fused silica column (30 m length × 0.32 mm i.d., 0.5 μm film thickness, Agilent J&W, USA). Sample injections of 2 μl were made using a 7683 Series autosampler (Agilent Technologies, Santa Clara, USA). Injections were carried out in a Split/Splitless injector kept at 250 °C. The split valve was opened for 30 s after the injection, with a purge flow of 44 ml min−1. Helium was used as the carrier gas at a constant flow of 1.5 ml min−1. The 1D GC oven temperature started at 40 °C and was then raised at 3 °C min−1 up to 240 °C and kept isothermal for 10 min. The 1D column outlet was connected to the 2D system by means of the MCS switching device. A deactivated fused silica column transferred 10% split from the MCS device to the FID in order to monitor the eluent from the 1D column. Additional permanent inlet and outlet capillary lines connected the MCS device to a mass flow controller and a pressure regulator module (EPC) which provided a constant inlet flow of 10 ml min−1 through the MCS device, thus allowing the heart-cut procedure. This counter-current flow was switched off during the transfer of a selected heart-cut in the 2D system. Heart-cut fractions were trapped in the first 50 cm of the 2D column across a cryo trap system (CTS, Gerstel) and then cooled at −150 °C with liquid nitrogen. After each heart-cut, the current flow was switched on and the CTS was heated to 250 °C at a rate of 20 °C s−1. The EPC module ensured programmed pressure at the head of the 2D column in order to maintain a constant flow of 1.5 ml min−1.
0/5000
จาก: -
เป็น: -
ผลลัพธ์ (ไทย) 1: [สำเนา]
คัดลอก!
แยกอุปกรณ์เป็น HP 6890 chromatograph (Agilent เทคโนโลยี ซานตาคลารา สหรัฐอเมริกา) พร้อมกับระบบเครื่องตรวจจับเปลวไฟ Ionisation (FID) เก็บไว้ที่ 250 ° C และขั้วขี้ผึ้ง DB นส่วน fused คอลัมน์ (30 × m ความยาว 0.32 มม.ประชาชน ความหนาฟิล์ม 0.5 μm, Agilent J และ W สหรัฐอเมริกา) ฉีดตัวอย่างของ 2 μl แปลงโดยใช้ autosampler เป็นชุด 7683 (Agilent เทคโนโลยี ซานตาคลารา สหรัฐอเมริกา) ฉีดได้ดำเนินการอัด แยก/Splitless ที่เก็บไว้ที่ 250 องศาเซลเซียส เปิดวาล์วแยกสำหรับ 30 s หลังจากการฉีด มีขั้นตอนการล้างข้อมูลของ min−1 44 ml มีใช้ฮีเลียมเป็นก๊าซผู้ขนส่งที่ไหลคง min−1 1.5 ml เริ่มต้นที่ 40 ° C อุณหภูมิเตาอบ GC 1D และขึ้นแล้วที่ min−1 3 ° C ถึง 240 ° C และ isothermal เก็บไว้สำหรับ 10 นาที ร้านคอลัมน์ 1D ถูกเชื่อมต่อกับระบบ 2D โดยใช้อุปกรณ์สลับเอ็มซี คอลัมน์ปิด fused นส่วนแยกจากอุปกรณ์เอ็มซีสลักเพื่อติดตาม eluent จากคอลัมน์ 1 D 10% แล้ว บรรทัดเพิ่มเติมถาวรร้านและทางเข้าของเส้นเลือดฝอยเชื่อมต่ออุปกรณ์เอ็มซีกับตัวควบคุมการไหลเชิงมวลและความดันควบคุมโมดูล (EPC) ซึ่งให้กระแสคงทางเข้าของของ min−1 10 ml ผ่านอุปกรณ์เอ็มซี จึง ช่วยให้กระบวนการตัดใจ กระแสทวนปัจจุบันนี้ได้ถูกปิดการตัดใจเลือกในระบบ 2D เศษตัดใจถูกติดอยู่ใน 50 cm แรกของคอลัมน์ 2D ข้ามระบบการดัก cryo (CTS, Gerstel) และระบายความร้อนด้วยที่ −150 ° C ด้วยไนโตรเจนเหลวแล้ว หลังจากแต่ละห้องตัด สลับในกระแสปัจจุบัน และ CTS ที่อุณหภูมิ 250 ° C ในอัตรา 20 ° C s−1 โม EPC มั่นใจโปรแกรมความดันที่หัวของคอลัมน์ 2D เพื่อรักษากระแสคงที่ของ min−1 1.5 ml
การแปล กรุณารอสักครู่..
ผลลัพธ์ (ไทย) 2:[สำเนา]
คัดลอก!
separation device was a HP 6890 chromatograph (Agilent Technologies, Santa Clara, USA) equipped with a Flame Ionisation Detector (FID) system kept at 250 °C and a polar DB-Wax fused silica column (30 m length × 0.32 mm i.d., 0.5 μm film thickness, Agilent J&W, USA). Sample injections of 2 μl were made using a 7683 Series autosampler (Agilent Technologies, Santa Clara, USA). Injections were carried out in a Split/Splitless injector kept at 250 °C. The split valve was opened for 30 s after the injection, with a purge flow of 44 ml min−1. Helium was used as the carrier gas at a constant flow of 1.5 ml min−1. The 1D GC oven temperature started at 40 °C and was then raised at 3 °C min−1 up to 240 °C and kept isothermal for 10 min. The 1D column outlet was connected to the 2D system by means of the MCS switching device. A deactivated fused silica column transferred 10% split from the MCS device to the FID in order to monitor the eluent from the 1D column. Additional permanent inlet and outlet capillary lines connected the MCS device to a mass flow controller and a pressure regulator module (EPC) which provided a constant inlet flow of 10 ml min−1 through the MCS device, thus allowing the heart-cut procedure. This counter-current flow was switched off during the transfer of a selected heart-cut in the 2D system. Heart-cut fractions were trapped in the first 50 cm of the 2D column across a cryo trap system (CTS, Gerstel) and then cooled at −150 °C with liquid nitrogen. After each heart-cut, the current flow was switched on and the CTS was heated to 250 °C at a rate of 20 °C s−1. The EPC module ensured programmed pressure at the head of the 2D column in order to maintain a constant flow of 1.5 ml min−1.
การแปล กรุณารอสักครู่..
ผลลัพธ์ (ไทย) 3:[สำเนา]
คัดลอก!
อุปกรณ์แยกเป็น HP 6890 Chromatograph ( Agilent Technologies , ซานตาคลารา , USA ) อุปกรณ์ตรวจจับเปลวไฟ ionisation ( FID ) ระบบไว้ที่ 250 องศา C และขี้ผึ้ง ดีบี ขั้วฟิวส์คอลัมน์ซิลิกา ( 30 เมตรความยาว× 0.32 มม. บัตร , 0.5 μเมตรความหนาฟิล์ม Agilent J & W , USA ) ตัวอย่างการฉีด 2 μผมได้ทําการใช้ 7683 autosampler ชุด ( Agilent Technologies , ซานตาคลารา สหรัฐอเมริกา )ฉีดอะมิโน Split / splitless ฉีดไว้ที่ 250 องศา แยกวาล์วเปิด 30 วินาทีหลังจากฉีดกับล้างไหล 44 มิลลิลิตรต่อนาที− 1 ฮีเลียมถูกใช้เป็นแก๊สตัวพาในการไหลคงที่ของ 1.5 มิลลิลิตรต่อนาที− 1 โค้ด 1D GC เตาอบที่อุณหภูมิเริ่มต้นที่ 40 ° C และก็ยกที่ 3 ° C มิน− 1 ถึง 240 ° C และเก็บอุณหภูมิสำหรับ 10 นาทีโค้ด 1D คอลัมน์เต้าเสียบถูกเชื่อมต่อกับระบบ 2D โดยวิธีการทางการเปลี่ยนอุปกรณ์ ปิดการใช้งานผสมคอลัมน์ซิลิกาโอน 10% แยกจาก MCS อุปกรณ์กับสลักเพื่อที่จะติดตามตัวจาก 1D คอลัมน์เพิ่มเติมถาวร ขาเข้า ขาออกฝอยเส้นเชื่อมต่อ MCS อุปกรณ์เพื่อควบคุมการไหลของมวลและความดัน Regulator Module ( EPC ) ซึ่งมีการไหลคงที่ของ 10 มิลลิลิตรต่อนาที− 1 ผ่านพิธีกร อุปกรณ์ จึงช่วยให้กระบวนการตัดใจ ในปัจจุบันนี้นับการปิดในช่วงการเลือกหัวใจตัดในระบบ 2 มิติควักหัวใจ เศษส่วน ถูกขังอยู่ในแรก 50 ซม. ของ 2D คอลัมน์ข้ามระบบกับดัก แช่แข็ง ( CTS อนเกอร์ ทล , ) แล้วที่− 150 ° C เย็นด้วยไนโตรเจนเหลว หลังจากที่แต่ละหัวใจตัดการไหลของกระแสสลับบนและ CTS ให้ความร้อน 250 ° C ในอัตรา 20 ° C − 1 โมดูลโปรแกรม EPC ทำให้แรงดันที่หัวของคอลัมน์ 2D เพื่อรักษาไหลคงที่ของ 15 มิลลิลิตรต่อนาที− 1
การแปล กรุณารอสักครู่..
 
ภาษาอื่น ๆ
การสนับสนุนเครื่องมือแปลภาษา: กรีก, กันนาดา, กาลิเชียน, คลิงออน, คอร์สิกา, คาซัค, คาตาลัน, คินยารวันดา, คีร์กิซ, คุชราต, จอร์เจีย, จีน, จีนดั้งเดิม, ชวา, ชิเชวา, ซามัว, ซีบัวโน, ซุนดา, ซูลู, ญี่ปุ่น, ดัตช์, ตรวจหาภาษา, ตุรกี, ทมิฬ, ทาจิก, ทาทาร์, นอร์เวย์, บอสเนีย, บัลแกเรีย, บาสก์, ปัญจาป, ฝรั่งเศส, พาชตู, ฟริเชียน, ฟินแลนด์, ฟิลิปปินส์, ภาษาอินโดนีเซี, มองโกเลีย, มัลทีส, มาซีโดเนีย, มาราฐี, มาลากาซี, มาลายาลัม, มาเลย์, ม้ง, ยิดดิช, ยูเครน, รัสเซีย, ละติน, ลักเซมเบิร์ก, ลัตเวีย, ลาว, ลิทัวเนีย, สวาฮิลี, สวีเดน, สิงหล, สินธี, สเปน, สโลวัก, สโลวีเนีย, อังกฤษ, อัมฮาริก, อาร์เซอร์ไบจัน, อาร์เมเนีย, อาหรับ, อิกโบ, อิตาลี, อุยกูร์, อุสเบกิสถาน, อูรดู, ฮังการี, ฮัวซา, ฮาวาย, ฮินดี, ฮีบรู, เกลิกสกอต, เกาหลี, เขมร, เคิร์ด, เช็ก, เซอร์เบียน, เซโซโท, เดนมาร์ก, เตลูกู, เติร์กเมน, เนปาล, เบงกอล, เบลารุส, เปอร์เซีย, เมารี, เมียนมา (พม่า), เยอรมัน, เวลส์, เวียดนาม, เอสเปอแรนโต, เอสโทเนีย, เฮติครีโอล, แอฟริกา, แอลเบเนีย, โคซา, โครเอเชีย, โชนา, โซมาลี, โปรตุเกส, โปแลนด์, โยรูบา, โรมาเนีย, โอเดีย (โอริยา), ไทย, ไอซ์แลนด์, ไอร์แลนด์, การแปลภาษา.

Copyright ©2025 I Love Translation. All reserved.

E-mail: