separation device was a HP 6890 chromatograph (Agilent Technologies, Santa Clara, USA) equipped with a Flame Ionisation Detector (FID) system kept at 250 °C and a polar DB-Wax fused silica column (30 m length × 0.32 mm i.d., 0.5 μm film thickness, Agilent J&W, USA). Sample injections of 2 μl were made using a 7683 Series autosampler (Agilent Technologies, Santa Clara, USA). Injections were carried out in a Split/Splitless injector kept at 250 °C. The split valve was opened for 30 s after the injection, with a purge flow of 44 ml min−1. Helium was used as the carrier gas at a constant flow of 1.5 ml min−1. The 1D GC oven temperature started at 40 °C and was then raised at 3 °C min−1 up to 240 °C and kept isothermal for 10 min. The 1D column outlet was connected to the 2D system by means of the MCS switching device. A deactivated fused silica column transferred 10% split from the MCS device to the FID in order to monitor the eluent from the 1D column. Additional permanent inlet and outlet capillary lines connected the MCS device to a mass flow controller and a pressure regulator module (EPC) which provided a constant inlet flow of 10 ml min−1 through the MCS device, thus allowing the heart-cut procedure. This counter-current flow was switched off during the transfer of a selected heart-cut in the 2D system. Heart-cut fractions were trapped in the first 50 cm of the 2D column across a cryo trap system (CTS, Gerstel) and then cooled at −150 °C with liquid nitrogen. After each heart-cut, the current flow was switched on and the CTS was heated to 250 °C at a rate of 20 °C s−1. The EPC module ensured programmed pressure at the head of the 2D column in order to maintain a constant flow of 1.5 ml min−1.
separation device was a HP 6890 chromatograph (Agilent Technologies, Santa Clara, USA) equipped with a Flame Ionisation Detector (FID) system kept at 250 °C and a polar DB-Wax fused silica column (30 m length × 0.32 mm i.d., 0.5 μm film thickness, Agilent J&W, USA). Sample injections of 2 μl were made using a 7683 Series autosampler (Agilent Technologies, Santa Clara, USA). Injections were carried out in a Split/Splitless injector kept at 250 °C. The split valve was opened for 30 s after the injection, with a purge flow of 44 ml min−1. Helium was used as the carrier gas at a constant flow of 1.5 ml min−1. The 1D GC oven temperature started at 40 °C and was then raised at 3 °C min−1 up to 240 °C and kept isothermal for 10 min. The 1D column outlet was connected to the 2D system by means of the MCS switching device. A deactivated fused silica column transferred 10% split from the MCS device to the FID in order to monitor the eluent from the 1D column. Additional permanent inlet and outlet capillary lines connected the MCS device to a mass flow controller and a pressure regulator module (EPC) which provided a constant inlet flow of 10 ml min−1 through the MCS device, thus allowing the heart-cut procedure. This counter-current flow was switched off during the transfer of a selected heart-cut in the 2D system. Heart-cut fractions were trapped in the first 50 cm of the 2D column across a cryo trap system (CTS, Gerstel) and then cooled at −150 °C with liquid nitrogen. After each heart-cut, the current flow was switched on and the CTS was heated to 250 °C at a rate of 20 °C s−1. The EPC module ensured programmed pressure at the head of the 2D column in order to maintain a constant flow of 1.5 ml min−1.
การแปล กรุณารอสักครู่..

อุปกรณ์แยกเป็น HP 6890 Chromatograph ( Agilent Technologies , ซานตาคลารา , USA ) อุปกรณ์ตรวจจับเปลวไฟ ionisation ( FID ) ระบบไว้ที่ 250 องศา C และขี้ผึ้ง ดีบี ขั้วฟิวส์คอลัมน์ซิลิกา ( 30 เมตรความยาว× 0.32 มม. บัตร , 0.5 μเมตรความหนาฟิล์ม Agilent J & W , USA ) ตัวอย่างการฉีด 2 μผมได้ทําการใช้ 7683 autosampler ชุด ( Agilent Technologies , ซานตาคลารา สหรัฐอเมริกา )ฉีดอะมิโน Split / splitless ฉีดไว้ที่ 250 องศา แยกวาล์วเปิด 30 วินาทีหลังจากฉีดกับล้างไหล 44 มิลลิลิตรต่อนาที− 1 ฮีเลียมถูกใช้เป็นแก๊สตัวพาในการไหลคงที่ของ 1.5 มิลลิลิตรต่อนาที− 1 โค้ด 1D GC เตาอบที่อุณหภูมิเริ่มต้นที่ 40 ° C และก็ยกที่ 3 ° C มิน− 1 ถึง 240 ° C และเก็บอุณหภูมิสำหรับ 10 นาทีโค้ด 1D คอลัมน์เต้าเสียบถูกเชื่อมต่อกับระบบ 2D โดยวิธีการทางการเปลี่ยนอุปกรณ์ ปิดการใช้งานผสมคอลัมน์ซิลิกาโอน 10% แยกจาก MCS อุปกรณ์กับสลักเพื่อที่จะติดตามตัวจาก 1D คอลัมน์เพิ่มเติมถาวร ขาเข้า ขาออกฝอยเส้นเชื่อมต่อ MCS อุปกรณ์เพื่อควบคุมการไหลของมวลและความดัน Regulator Module ( EPC ) ซึ่งมีการไหลคงที่ของ 10 มิลลิลิตรต่อนาที− 1 ผ่านพิธีกร อุปกรณ์ จึงช่วยให้กระบวนการตัดใจ ในปัจจุบันนี้นับการปิดในช่วงการเลือกหัวใจตัดในระบบ 2 มิติควักหัวใจ เศษส่วน ถูกขังอยู่ในแรก 50 ซม. ของ 2D คอลัมน์ข้ามระบบกับดัก แช่แข็ง ( CTS อนเกอร์ ทล , ) แล้วที่− 150 ° C เย็นด้วยไนโตรเจนเหลว หลังจากที่แต่ละหัวใจตัดการไหลของกระแสสลับบนและ CTS ให้ความร้อน 250 ° C ในอัตรา 20 ° C − 1 โมดูลโปรแกรม EPC ทำให้แรงดันที่หัวของคอลัมน์ 2D เพื่อรักษาไหลคงที่ของ 15 มิลลิลิตรต่อนาที− 1
การแปล กรุณารอสักครู่..
