When the DRS4 recording is stopped, the currents flowing in and out of the DRS4 chip suddenly change, causing a visible baseline transient in the transparent mode signal. If needed, this baseline transient can be measured and calibrated, since it is constant for a given baseline offset and DRS4 sampling speed.