Regarding there are requirement for storage temperature of finished product on IFU are below 25° C and in the Finished Good warehouse at PSI were controlled temperature by install air conditioning system and setting temperature at 25°C . However there are not monitoring temperature in the warehouse. Prior to monitoring temperature , the reference location to monitoring temperature should be review and selected from the hottest spot .
The temperature mapping should be checked and review the location in order to find out the hottest spot