For the calculation of the overall measurement uncertainty,
the calibration accuracy error, the frequency response
error of the instrument (which was calculated at
1%), and the nonuniform field error (which was calculated
at 0.8%) were considered. The expanded uncertainty was
calculated at 5.5% for the measurements of B and 6.3% for
the percentage of the 1998 ICNIRP reference level measurements
[10].