For ความปลอดภัย reasons, the [OFF] button is always active to allow รังสีเอ็กซ์เรย์ radiation to be switched off at any time.
Automatic centering can also be started at any desired high voltage (e.g. if this high voltage is being used exclusively).
High voltage and tube current must be adjusted in line with the inspection activity being performed.
To do this, they must be perfectly coordinated with one another using the image parameters "Sharpness", "Contrast" and "Noise" (see x|act base software).
High voltage and tube current can be preset before รังสีเอ็กซ์เรย์ generation and adjusted further during it.
Start with low beam power (e.g. 60 kV / 20 µA).
At beam powers in excess of approx. 4 W, the focus is widened to avoid penetration damage to the target.
The size of the focus is always optimally adjusted to the physical toughness of the target.