SAT (Site Acceptance Test)
The SAT (site acceptance test) will include the following tests:
Finishing Visual check
Main components visual check
Internal box pressure and ventilation setting
Utilities functionality and setting check
Functionality/Interlocks Verification (Mechanical & Software)
Hot test for dispensing systems (activity source supplied by Customer)
Dose calibrator verification (activity source supplied by Customer)
Safety devices and interlocks check
Operator’s training
Delivery of the performed FAT protocol and documentation package including:
Performed FAT Protocol
Maintenance and User’s manual
Recommended spare parts list
Certificate of compliance
As built technical drawings (electrical, mechanical, pneumatic & process schemes)
Materials certificates/data sheets
Main equipments data sheets
Instruments calibration certificates
Welding Processes qualification