Monitor Measurements
We analyzed 124 blood specimen pairs using the
monitor and the laboratory analyzer. The INR values showed
correlations of R2 = 0.8596 between plasma samples (BCT
coagulometer) and capillary blood samples (AvoSure PT
PRO) and R2 = 0.7387 between plasma samples (BCT coagulometer
and AvoSure PT PRO)