The instrumental parameters of the ICP were optimized based
on plasma robustness as reported for the analysis of milk (Silva
et al., 2004), dietetic sweeteners (Sousa et al., 2011) and also
biologic fluids (Bortoleto et al., 2010). The concept of robust plasma
is related to characteristics such as stability and high temperatures,
which are adequate for complex matrices, allowing the decomposition
of organic molecules and minimizing interferences due to
matrix effects (Brenner and Zander, 2000). The parameter
robustness is obtained from the ratio of atomic (285.213 nm)
and ionic (280.270 nm) emission lines of Mg and the plasma is
considered robust when the Mg II/Mg I ratio is equal to or higher
than 8. In this context, the nebulizer gas flow rate (NFR) was
evaluated by fixing the plasma power, the auxiliary gas flow rate
and by varying the NFR. After fixing the NFR on the most adequate
condition and using the same auxiliary gas flow rate, the plasma
power was varied from 1200 to 1500 W.
พารามิเตอร์ที่เหมาะสมของเครื่องมือ ICP และความทนทานตาม
รายงานเพื่อการวิเคราะห์ของนม ( ซิลวา
et al . , 2004 ) , สารให้ความหวานอาหาร ( ซูซา et al . , 2011 ) และยัง
ของเหลวชีวภาพ ( bortoleto et al . , 2010 ) แนวคิดของ
พลาสมาที่แข็งแกร่งเกี่ยวข้องกับคุณลักษณะ เช่น ความมั่นคงและอุณหภูมิสูง ซึ่งเพียงพอสำหรับ
เมทริกซ์ที่ซับซ้อนทำให้การสลายตัวของโมเลกุลอินทรีย์และลด
ผลการแทรกแซงจากเมทริกซ์ ( เบรนเนอร์กับแซนเดอร์ , 2000 ) พารามิเตอร์
ทนทานได้จากอัตราส่วนของอะตอม ( 285.213 nm )
) และไอออน ( 280.270 nm ) หลักล้านของมก. และพลาสม่า
ถือว่าแข็งแกร่งเมื่อมิลลิกรัม / MG II ผมเท่ากับหรือสูงกว่า
กว่า 8 ในบริบทนี้ อัตราการไหลของแก๊ส , nebulizer ( nfr ) คือ
ประเมิน โดยการเสริมพลังพลาสม่า , อัตราการไหลของแก๊ส
และเปลี่ยนแปลง nfr . หลังจากแก้ไข nfr บนเงื่อนไขที่เพียงพอ
ที่สุดและใช้อัตราการไหลของแก๊สแบบเสริมพลังพลาสม่า
มีค่าตั้งแต่ 1 , 200 ถึง 1 , 500 W
การแปล กรุณารอสักครู่..