Dear Wachira,
thanks for your information.
Let’s do it in brief:
MELAtronic Err32; That Err appears only if no electricity is detected or less than 200 V.
Often the user terminate the program by main switch. If they switch on again that message will appear in order to inform them about electricity cut-off. The feed pump is not connected with that err message.
Pls check again i.e. in the diagnosis menu.
Vacuklav 23 B+
We will add a new thermal fuse to next delivery. However pls check what was the reason. It is a thermal problem of steam generator, may be by overloading. You can also check the feeding rate of feed pump, however in a new autoclave it shouldn’t be that case. Any inlet filter blocked or dirty measurement sensor housing ?