Maintenance and adjustment
Warning
When executing maintenance sequences, be sure to unplug the instrument first to avoid electric shock.
•For models with no power plug, turn off the breaker to which the power cord is connected.
Caution
When executing maintenance sequences, make sure that the temperature inside the chamber falls down sufficiently.
•Failure to do so can cause burns to the user.
CLEANING AND DISINFECTING THE MAIN UNIT
When the outer panel or the inside the chamber is contaminated, perform cleaning and disinfecting in accordance with the following method:
(1) Wipe exterior or interior of the unit with a soft cloth dampened in mild detergent.
(2) Thoroughly wipe off all traces of the detergent with a tightly squeezed damp cloth.
(1) Disinfect the unit using ethanol.
REPLACING BELLOWS
When the Bellows indication lamp lights or flashes in red or green, replace bellows of air release valve.
*The Bellows indication lamp may also light or flash, when the volume of material to be sterilized is too much as compared to the size of the chamber. Check the volume of sterilizing material.