If, instead, you want to measure in direct readings (response factor = 1) for one specific compound
with maximum accuracy over a wide range of concentrations, perform a multipoint calibration
with the specific compound. Up to 9 span points (plus zero) can be entered for each
detector. The use of a response curve is thus unnecessary as the detector is already reading the
direct PPM for that specific compound.