Fluorine was determined via the rot

Fluorine was determined via the rotational molecular absorption line of aluminum mono fluoride (AlF)
generated in C2H2/N2O flame at 227.4613 nm using a high-resolution continuum source flame atomic
absorption spectrophotometer (HR-CS-FAAS).
The effects of AlF wavelength, burner height, fuel rate (C2H2/N2O) and amount of Al on the accuracy, precision and sensitivity were investigated and optimized.
The Al–F absorption band at 227.4613 nm was found to be the most suitable analytical line with respect
to sensitivity and spectral interferences.
Maximum sensitivity and a good linearity were obtained in
acetylene-nitrous oxide flame at a flow rate of 210 L h−1 and a burner height of 8 mm using 3000 mg L−1
of Al for 10–1000 mg L−1of F.
The accuracy and precision of the method were tested by analyzing spiked
samples and waste water certified reference material.
The results were in good agreement with the certified
and spiked amounts as well as the precision of several days during this study was satisfactory
(RSD < 10%).
The limit of detection and characteristic concentration of the method were 5.5 mg L−1 and
72.8 mg L−1, respectively.
Finally, the fluorine concentrations in several toothpaste samples were determined.
The results found and given by the producers were not significantly different.
The method was simple, fast, accurate and sensitive.
0/5000
จาก: -
เป็น: -
ผลลัพธ์ (ไทย) 1: [สำเนา]
คัดลอก!
ฟลูออรีนกำหนดผ่านสายดูดซึมโมเลกุลในการหมุนของโมโนฟลูออไรด์อลูมิเนียม (AlF)สร้างขึ้นในเปลวไฟ C2H2/N2O ที่ 227.4613 nm โดยใช้แหล่งมีความละเอียดสูงสมิติเฟลมอะตอมดูดซึมเครื่องทดสอบกรดด่าง (HR CS FAAS) ผลกระทบของ AlF ความยาวคลื่น ความสูงของเตา อัตราน้ำมันเชื้อเพลิง (C2H2 N2O) และจำนวนอัลบนความถูกต้อง ความแม่นยำ และความไวถูกสอบสวน และปรับให้เหมาะสมวงดูดซึมอัล – F ที่ 227.4613 nm พบเป็น บรรทัดวิเคราะห์เหมาะสมที่สุด ด้วยความเคารพความไวและ interferences สเปกตรัม ความไวสูงสุดและแบบดอกไม้ที่ดีได้รับในเปลวไฟกับอะเซทิลีนไนตรัสออกไซด์ในอัตราการไหลของ h−1 210 L และเขียนความสูง 8 มม.ใช้ 3000 mg L−1ของอัลสำหรับเอฟ L−1of 10 – 1000 มิลลิกรัม ความถูกต้องและความแม่นยำของวิธีทดสอบ โดยการวิเคราะห์ spikedตัวอย่างน้ำเสียรับรองเอกสารอ้างอิง ผลลัพธ์ได้ในข้อตกลงที่ดีกับการได้รับการรับรองและยอดเงินที่ถูกแทงเป็นความแม่นยำในการศึกษานี้หลายวันก็พอ(RSD < 10%) จำนวนเข้มข้นวิธีการตรวจสอบและลักษณะได้ 5.5 มิลลิกรัม L−1 และ72.8 มิลลิกรัม L−1 ตามลำดับ สุดท้าย ความเข้มข้นฟลูออรีนในตัวอย่างยาสีฟันหลายถูกกำหนดผลการค้นพบ และกำหนด โดยผู้ผลิตที่ไม่แตกต่างกันอย่างมีนัยสำคัญ วิธีง่าย รวดเร็ว ถูกต้อง และสำคัญ
การแปล กรุณารอสักครู่..
ผลลัพธ์ (ไทย) 2:[สำเนา]
คัดลอก!
Fluorine was determined via the rotational molecular absorption line of aluminum mono fluoride (AlF)
generated in C2H2/N2O flame at 227.4613 nm using a high-resolution continuum source flame atomic
absorption spectrophotometer (HR-CS-FAAS).
The effects of AlF wavelength, burner height, fuel rate (C2H2/N2O) and amount of Al on the accuracy, precision and sensitivity were investigated and optimized.
The Al–F absorption band at 227.4613 nm was found to be the most suitable analytical line with respect
to sensitivity and spectral interferences.
Maximum sensitivity and a good linearity were obtained in
acetylene-nitrous oxide flame at a flow rate of 210 L h−1 and a burner height of 8 mm using 3000 mg L−1
of Al for 10–1000 mg L−1of F.
The accuracy and precision of the method were tested by analyzing spiked
samples and waste water certified reference material.
The results were in good agreement with the certified
and spiked amounts as well as the precision of several days during this study was satisfactory
(RSD < 10%).
The limit of detection and characteristic concentration of the method were 5.5 mg L−1 and
72.8 mg L−1, respectively.
Finally, the fluorine concentrations in several toothpaste samples were determined.
The results found and given by the producers were not significantly different.
The method was simple, fast, accurate and sensitive.
การแปล กรุณารอสักครู่..
ผลลัพธ์ (ไทย) 3:[สำเนา]
คัดลอก!
ฟลูออรีนถูกกำหนดผ่านการดูดซับโมเลกุลสายอลูมิเนียมโมโนฟลูออไรด์ ( ALF )
สร้างขึ้นใน c2h2 / N2O เปลวไฟที่ 227.4613 nm ใช้ความละเอียดสูงต่อเนื่องแหล่งเฟลมอะตอม
วัสดุดูดซับ ( hr-cs-faas )
ผลของอัลฟ์ความยาวคลื่นสูง Burner , อัตราเชื้อเพลิง ( c2h2 / N2O ) และปริมาณของอัลเกี่ยวกับความถูกต้องความแม่นยำและความไว ) ได้ศึกษาและปรับให้เหมาะสม อัล–
F การดูดกลืนที่ 227.4613 nm ได้เหมาะสมที่สุดเชิงเส้น ด้วยความเคารพ
ความไวสเปกตรัมและการแทรกแซง .
ความไวสูงสุดและได้รับกระแสที่ดีใน
อะเซทิลีนเปลวไฟก๊าซไนตรัสออกไซด์ที่อัตราการ ไหลของ 210 L H − 1 และเครื่องเขียน ความสูง 8 มม. ใช้ 3000 mg L − 1
ของ Al 10 – 1000 mg L − 1of F .
ความถูกต้องและความแม่นยำของวิธีการทดสอบโดยวิเคราะห์ตัวอย่างน้ำและถูกแทง
ของเสียวัสดุอ้างอิงที่ได้รับการรับรอง .
ผลสอดคล้องกับการรับรอง
และถูกแทงยอด รวมทั้งความแม่นยําของหลายวันในช่วงการศึกษานี้ น่าพอใจ
( RSD < 10% )
ขีดจำกัดของการตรวจหาปริมาณและลักษณะของวิธี 5.5 mg L − 1
6 หมวด L − 1 มิลลิกรัมต่อลิตร ตามลำดับ
ในที่สุด ฟลูออรีน ความเข้มข้นของตัวอย่างยาสีฟันหลายตัวอย่าง
พบและได้รับโดยผู้ผลิตไม่แตกต่างกันอย่างมีนัยสำคัญทางสถิติ
วิธีที่ง่าย รวดเร็ว ถูกต้อง และอ่อนไหว
การแปล กรุณารอสักครู่..
 
ภาษาอื่น ๆ
การสนับสนุนเครื่องมือแปลภาษา: กรีก, กันนาดา, กาลิเชียน, คลิงออน, คอร์สิกา, คาซัค, คาตาลัน, คินยารวันดา, คีร์กิซ, คุชราต, จอร์เจีย, จีน, จีนดั้งเดิม, ชวา, ชิเชวา, ซามัว, ซีบัวโน, ซุนดา, ซูลู, ญี่ปุ่น, ดัตช์, ตรวจหาภาษา, ตุรกี, ทมิฬ, ทาจิก, ทาทาร์, นอร์เวย์, บอสเนีย, บัลแกเรีย, บาสก์, ปัญจาป, ฝรั่งเศส, พาชตู, ฟริเชียน, ฟินแลนด์, ฟิลิปปินส์, ภาษาอินโดนีเซี, มองโกเลีย, มัลทีส, มาซีโดเนีย, มาราฐี, มาลากาซี, มาลายาลัม, มาเลย์, ม้ง, ยิดดิช, ยูเครน, รัสเซีย, ละติน, ลักเซมเบิร์ก, ลัตเวีย, ลาว, ลิทัวเนีย, สวาฮิลี, สวีเดน, สิงหล, สินธี, สเปน, สโลวัก, สโลวีเนีย, อังกฤษ, อัมฮาริก, อาร์เซอร์ไบจัน, อาร์เมเนีย, อาหรับ, อิกโบ, อิตาลี, อุยกูร์, อุสเบกิสถาน, อูรดู, ฮังการี, ฮัวซา, ฮาวาย, ฮินดี, ฮีบรู, เกลิกสกอต, เกาหลี, เขมร, เคิร์ด, เช็ก, เซอร์เบียน, เซโซโท, เดนมาร์ก, เตลูกู, เติร์กเมน, เนปาล, เบงกอล, เบลารุส, เปอร์เซีย, เมารี, เมียนมา (พม่า), เยอรมัน, เวลส์, เวียดนาม, เอสเปอแรนโต, เอสโทเนีย, เฮติครีโอล, แอฟริกา, แอลเบเนีย, โคซา, โครเอเชีย, โชนา, โซมาลี, โปรตุเกส, โปแลนด์, โยรูบา, โรมาเนีย, โอเดีย (โอริยา), ไทย, ไอซ์แลนด์, ไอร์แลนด์, การแปลภาษา.

Copyright ©2024 I Love Translation. All reserved.

E-mail: