Check that the installation site is physically acceptable as follows:
• With the customer, evaluate site access for delivery. Consider the availability of a loading
dock, elevators, ramps, and the width of passageways and doors.
• Verify that there is adequate space available for system uncrating and installation.
• At each location where the system will be operated, verify that there is enough space for the
system and any external peripherals.
• Check adjacent rooms for activity and types of equipment that might negatively affect the
ultrasound system.