CTS’s parts are designed to be compatible
and interchangeable with the OEM parts.
1. Power down the PDA detector.
2. Remove the power cord.
3. Wait 15 minutes for the detector
lamp to cool before handling the
lamp.
4. Remove the optics cover to access
the lamp.
5. Disconnect the lamp power
connector.
6. Unscrew the two mounting screws.
7. Grip the metal base and pull the
lamp out.
8. Remove the mounting screws from
the old lamp.
9. Carefully unpack and hold the
replacement lamp by its base only.
10. Orient the lamp so the two holes in
the base align with the mounting pins
and screws in the optics bench.
11. Insert the lamp and secure it with the
two mounting screws.
12. Reconnect the lamp power
connector.
13. Close detector housing.
14. Reconnect the power cord and power
up PDA detector.
15. Record initial lamp intensity in the
instrument log.