11. Report
11.1 If the range of three successive observations of the
aniline point temperature is not greater than 0.1°C (0.2°F) for
light-colored samples or 0.2°C (0.4°F) for dark samples, report
the average temperature of these observations, corrected for
thermometer calibration errors, to the nearest 0.05°C (0.1°F) as
the aniline point.
11.2 If such a range is not obtained after five observations,
repeat the test using fresh quantities of aniline and sample in a
clean, dry apparatus, and if consecutive temperature observations
show a progressive change, or if the range of observations
is greater than the repeatability given in 12.1, report the test
method as being inapplicable