6.33 Suction Aspirator.
6.33.1 An electrically powered suction aspirator system shall be furnished.
6.33.2 The vacuum control, vacuum indicator, and collection bottle or bag shall be located so that it can be operated
from the primary patient care position.
6.33.4 The suction pump shall be located in an area that is accessible and insulated from the patient compartment.
6.33.5 The pump, when permanently mounted, shall be vented to the vehicle’s exterior.
6.33.6 A vacuum control and a shut-off valve, or combination thereof, shall be provided to adjust vacuum levels.
6.33.7 A vacuum indicator gauge graduated at least every 100 mm Hg and a minimum total range of 0 to 760 mm
Hg shall be provided