On Friday night we ran the System Calibration and it passed successfully.
On Saturday, we ran the 8x8 PPJ and it passed successfully.
We also installed a dry air hose near to the chiller black tubing hose where the frost was building up.
This was able to successfully eliminate the frost build up on the black tubing hose.
Exatron is also working on additional solutions to the frost build up issue.
When we left on Saturday we left the 5x5 kit running at room temperature.
When you arrive on Sunday, there will still be a few units left to run for the room PPJ.
Would it be possible for you please to run the 5x5 Hot & 5x5 cold PPG?
If you are able to do this, could you please take a screen capture of the hot & cold PPJ runs and put them in the PPJ Counts folder on the desk top?
Thank you very much for your assistance to John England and myself during the buyoff of the Exatron 903 Tri-Temp Handler.