6.2.2 A dynamic mode of titrant addition shall be used.
During the titration, the speed and volume of the addition shall vary depending on the rate of change of the system.
The recommended maximum volume increment is 0.5 mL and the recommended minimum volume increment is 0.05 mL.
6.2.3 Graduated Cylinder—50 mL, or dispensing device capable of delivering 50 6 0.5 mL.
6.2.4 Pipette—2.0 mL, Class A.
6.2.5 Titration Beaker—250 mL, 125 mL, or suitable capacity, made of borosilicate glass or other suitable material.
6.2.2 A dynamic mode of titrant addition shall be used. During the titration, the speed and volume of the addition shall vary depending on the rate of change of the system. The recommended maximum volume increment is 0.5 mL and the recommended minimum volume increment is 0.05 mL.6.2.3 Graduated Cylinder—50 mL, or dispensing device capable of delivering 50 6 0.5 mL.6.2.4 Pipette—2.0 mL, Class A.6.2.5 Titration Beaker—250 mL, 125 mL, or suitable capacity, made of borosilicate glass or other suitable material.