These machines are based on a number of force transducers, individually calibrated in a force standard machine and then loaded in parallel. The generated force is calculated as the sum of the forces being measured by the individual transducers. For this type of machine, the uncertainty contributions that need to be considered will include, but are not limited to, the following: