We included 50 patients in whom we used 105 sensors. Five patients from whom we could
not collect the predefined minimum number of four consecutive comparative blood draws were excluded from the point accuracy analysis.
Therefore, we had 929 comparative samples
from 100 sensors in 45 patients (11 (7 to 28) samples per patient) during 4,639 hours (46 (27to 134) hours per patient and 46 (21 to 69) hours per sensor) for the accuracy analysis.