The results presented in this paper are obtained by
measuring electromagnetic emissions from electrosurgery and
radiosurgery generators (all CE marked and available on the
market) in a laboratory, using a test setup very similar to the
one presented in the 60601-2-2 [5], in order to asses radiated
electromagnetic emissions. The measurements were made
using an Agilent E4404B ESA-E Series Spectrum Analyzer
with a frequency bandwidth of 9 kHz - 6.7 GHz and a
Schwarzbeck - EFS 9218-9218182 Active biconical antenna
with a frequency bandwidth of 9 kHz – 300 MHz, in vertical
polarization. The measurements were made between 10 kHz
and 30 MHz, using a 10 kHz resolution bandwidth, a 30 dB
attenuation at the spectrum analyzer and the reference level
was set at 20 dBm. The Max Hold function of the spectrum
analyzer was used in view of obtaining the maximum values
for the electric field.