By comparison, the worker flame channels milled directly into the monolithic housing of the SS mFP make for a much stronger device that is able to withstand routine aggressive handling without problems. Of note, no breakage has occurred in regular operation of the SS mFPD for over a year,whereas the quartz tube prototype normally required replacement every few weeks.