Plating procedure
Check stylus tip angle daily and adjust if necessary. (Use vacuum to hold microscope cover slip against face of stylus tip; if cover slip plane is parallel at about l mm from surface of platform, tip is properly oriented). Liquids are moved through system by vacuum. Clean stylus tip by rinsing for 1 s with sodium hypochlorite solution followed by sterile dilution water for 1 s before sample introduction. This rinse procedure between processing of each sample minimizes cross-contamination. After rinsing, draw sample into tip of Teflon tubing by vacuum applied to 2-way valve. When tubing and syringe are filled with sample, close valve attached to syringe. Place agar plate on platform, place stylus tip on agar surface, and start motor. During inoculation, label petri plate lid. After agar has been inoculated, stylus lifts from agar surface and spiral plater automatically stops. Remove inoculated plate from platform and cover it. Move stylus back to starting position. Vacuum-rinse system with hypochlorite and water, and then introduce new sample. Invert plates and promptly place them in incubator for 48 ± 3 h at 35 ± 1°C.