Can see graph of readings, menu 4, view graphs, (for XT software, press ]), can scroll through but normally keep an eye on photosynthesis and conductance and check on stability, etc.
Can check on recorded points by going to menu 1, view file (F2), QuikPik Config (F1) (first time, afterwards can RePlot Graph), select light curve.
Also while light curve is running, there is time (it usually take 20-30 min) to look for the next plant or do other measurements, such as leaf width, etc, just make sure battery level is OK or that you can hear the warning beeps if it becomes low.
When curve is finished, the star by menu 2 will be gone (a star indicates an auto-program is running), then check the graph of recorded points (see step 21), if the graph is OK and you are done with the leaf, then menu 1, close file (f3).
Now go to sensor head, while holding the leaf in place, open the cuvette and mark the leaf coverage of the cuvette, using a thin sharpie permanent marker.
Cut leaf off cleanly at stem, using a razor blade and put in a ziplock bag, keep the bag cool and out of direct light (in a cooler) until you can put in a leaf area meter.
If another leaf is being measured, go back to 10, otherwise put the instrument in sleep mode, download data or turn off.
- See more at: http://prometheuswiki.publish.csiro.au/tiki-print.php?page=GAS%20EXCHANGE%20PROTOCOL%20FOR%20LI-COR%206400#sthash.pcjwX8fE.dpuf
Can see graph of readings, menu 4, view graphs, (for XT software, press ]), can scroll through but normally keep an eye on photosynthesis and conductance and check on stability, etc.Can check on recorded points by going to menu 1, view file (F2), QuikPik Config (F1) (first time, afterwards can RePlot Graph), select light curve.Also while light curve is running, there is time (it usually take 20-30 min) to look for the next plant or do other measurements, such as leaf width, etc, just make sure battery level is OK or that you can hear the warning beeps if it becomes low.When curve is finished, the star by menu 2 will be gone (a star indicates an auto-program is running), then check the graph of recorded points (see step 21), if the graph is OK and you are done with the leaf, then menu 1, close file (f3).Now go to sensor head, while holding the leaf in place, open the cuvette and mark the leaf coverage of the cuvette, using a thin sharpie permanent marker.Cut leaf off cleanly at stem, using a razor blade and put in a ziplock bag, keep the bag cool and out of direct light (in a cooler) until you can put in a leaf area meter.If another leaf is being measured, go back to 10, otherwise put the instrument in sleep mode, download data or turn off.- See more at: http://prometheuswiki.publish.csiro.au/tiki-print.php?page=GAS%20EXCHANGE%20PROTOCOL%20FOR%20LI-COR%206400#sthash.pcjwX8fE.dpuf