*Found water flush out in the test chamber all the time and the status on the screen showed “filling tank”.
Check in the diagnostic screen and found VE(2) = +5 VDC.
The IO module and water sensor working correctly.
Open the tank and found that the float ball full of water.
Suggest Petpak to order new float ball (P/N PPT3-2-71-SVC - KIT, FLOAT REPL FOR PPT3-6-13 AGR SERVICEABLE PART) + Seal (P/N 2013925 - O-RING, 4.000""ID X 4.125""OD) and schedule for service visit or order new tank (P/N PPT3-6-13 - *ASSY, CYLINDER WITH FLOAT SENSOR).
"