As tabulated in Table 1, a total of eighteen sample coils which includ การแปล - As tabulated in Table 1, a total of eighteen sample coils which includ ไทย วิธีการพูด

As tabulated in Table 1, a total of

As tabulated in Table 1, a total of eighteen sample coils which includes plain, louver, and semi-dimple VG. The detailed dimension and the photo of the semi-dimple VG is schematically shown in Fig. 1 (a) whereas the louver fin is depicted in Fig. 1 (b). Notice that the fin thickness (D f), collar diameter ( D c), transverse pitch (Pt), and longitudinal pitch ( Pl) for all the test samples are 0.11 mm, 7.5 mm, 21 mm, and 18.2 mm, respectively. The corresponding fin pitch (Fp) ranges from 1.6–2.0 mm and the number of tube row (N)spans from 1 to 4 as shown in Table 1. Detailed construction of the circuitry arrangement is identical to those by Wang et al.[27] . The experiments are conducted in an open wind tunnel as shown in Fig. 2. The ambient air flow was forced across the test section by means of a 5.6 kW centrifugal fan with an inverter. To avoid and minimize the effect of flow maldistribution in the experiments, an air straightener-equalizer and a mixer were provided. The inlet and the exit temperatures across the sample coil were measured by two T-type thermocouple meshes. The inlet measuring mesh consists of twelve thermocouples while the outlet mesh contains 36 thermocouples. These data signals were individually recorded and then averaged. During the isothermal test, the variance of these thermocouples was within 0.2C. In addition, all the thermocouples were pre-calibrated by a quartz thermometer with 0.01C precision.
The pressure drop of the test coil was detected by a precision differential pressure transducer, reading to 0.1 Pa. The air flow measuring station was a multiple nozzle code tester based on the ASHRAE 41.2 standard [28] . The working medium in the tube side was hot water. The inlet water temperature was controlled by a thermostat reservoir having an adjustable capacity up to 25 kW. Both the inlet and outlet temperatures were measured by two precalibrated RTDs (Resistance temperature device, Pt-100). Their accuracy was within 0.05 C. The water volumetric flow rate is detected by a magnetic flow meter with 0.002 L/s resolution.
0/5000
จาก: -
เป็น: -
ผลลัพธ์ (ไทย) 1: [สำเนา]
คัดลอก!
As tabulated in Table 1, a total of eighteen sample coils which includes plain, louver, and semi-dimple VG. The detailed dimension and the photo of the semi-dimple VG is schematically shown in Fig. 1 (a) whereas the louver fin is depicted in Fig. 1 (b). Notice that the fin thickness (D f), collar diameter ( D c), transverse pitch (Pt), and longitudinal pitch ( Pl) for all the test samples are 0.11 mm, 7.5 mm, 21 mm, and 18.2 mm, respectively. The corresponding fin pitch (Fp) ranges from 1.6–2.0 mm and the number of tube row (N)spans from 1 to 4 as shown in Table 1. Detailed construction of the circuitry arrangement is identical to those by Wang et al.[27] . The experiments are conducted in an open wind tunnel as shown in Fig. 2. The ambient air flow was forced across the test section by means of a 5.6 kW centrifugal fan with an inverter. To avoid and minimize the effect of flow maldistribution in the experiments, an air straightener-equalizer and a mixer were provided. The inlet and the exit temperatures across the sample coil were measured by two T-type thermocouple meshes. The inlet measuring mesh consists of twelve thermocouples while the outlet mesh contains 36 thermocouples. These data signals were individually recorded and then averaged. During the isothermal test, the variance of these thermocouples was within 0.2C. In addition, all the thermocouples were pre-calibrated by a quartz thermometer with 0.01C precision.The pressure drop of the test coil was detected by a precision differential pressure transducer, reading to 0.1 Pa. The air flow measuring station was a multiple nozzle code tester based on the ASHRAE 41.2 standard [28] . The working medium in the tube side was hot water. The inlet water temperature was controlled by a thermostat reservoir having an adjustable capacity up to 25 kW. Both the inlet and outlet temperatures were measured by two precalibrated RTDs (Resistance temperature device, Pt-100). Their accuracy was within 0.05 C. The water volumetric flow rate is detected by a magnetic flow meter with 0.002 L/s resolution.
การแปล กรุณารอสักครู่..
ผลลัพธ์ (ไทย) 2:[สำเนา]
คัดลอก!
As tabulated in Table 1, a total of eighteen sample coils which includes plain, louver, and semi-dimple VG. The detailed dimension and the photo of the semi-dimple VG is schematically shown in Fig. 1 (a) whereas the louver fin is depicted in Fig. 1 (b). Notice that the fin thickness (D f), collar diameter ( D c), transverse pitch (Pt), and longitudinal pitch ( Pl) for all the test samples are 0.11 mm, 7.5 mm, 21 mm, and 18.2 mm, respectively. The corresponding fin pitch (Fp) ranges from 1.6–2.0 mm and the number of tube row (N)spans from 1 to 4 as shown in Table 1. Detailed construction of the circuitry arrangement is identical to those by Wang et al.[27] . The experiments are conducted in an open wind tunnel as shown in Fig. 2. The ambient air flow was forced across the test section by means of a 5.6 kW centrifugal fan with an inverter. To avoid and minimize the effect of flow maldistribution in the experiments, an air straightener-equalizer and a mixer were provided. The inlet and the exit temperatures across the sample coil were measured by two T-type thermocouple meshes. The inlet measuring mesh consists of twelve thermocouples while the outlet mesh contains 36 thermocouples. These data signals were individually recorded and then averaged. During the isothermal test, the variance of these thermocouples was within 0.2C. In addition, all the thermocouples were pre-calibrated by a quartz thermometer with 0.01C precision.
The pressure drop of the test coil was detected by a precision differential pressure transducer, reading to 0.1 Pa. The air flow measuring station was a multiple nozzle code tester based on the ASHRAE 41.2 standard [28] . The working medium in the tube side was hot water. The inlet water temperature was controlled by a thermostat reservoir having an adjustable capacity up to 25 kW. Both the inlet and outlet temperatures were measured by two precalibrated RTDs (Resistance temperature device, Pt-100). Their accuracy was within 0.05 C. The water volumetric flow rate is detected by a magnetic flow meter with 0.002 L/s resolution.
การแปล กรุณารอสักครู่..
ผลลัพธ์ (ไทย) 3:[สำเนา]
คัดลอก!
ตารางที่ 1 เป็นตาราง มีทั้งหมด 18 ตัวอย่างขดลวดซึ่งรวมถึงธรรมดา , บานเกล็ด , และกึ่งลักยิ้ม วีจี มิติรายละเอียดและรูปถ่ายของกึ่งลักยิ้ม VG คือแผนผังแสดงในรูปที่ 1 ( a ) ส่วนบานครีบเป็นภาพในรูปที่ 1 ( B ) สังเกตว่าครีบหนา ( D F ) , คอเสา ( D C ) , สนามตามขวาง ( PT ) และระดับเสียงตามยาว ( PL ) สำหรับตัวอย่างทดสอบทั้งหมดเป็น 011 มม. 7.5 มม. 21 มม. และ 18.2 มิลลิเมตร ตามลำดับ ระยะห่างครีบที่สอดคล้องกัน ( FP ) ช่วงจาก 1.6 และ 2.0 มม. และจำนวนแถวท่อ ( N ) ขยายจาก 1 ดังแสดงในตารางที่ 1 รายละเอียดการก่อสร้างของ วงจรการจัดเหมือนที่วัง et al . [ 27 ] การทดลองมีวัตถุประสงค์ในการเปิดอุโมงค์ลม ดังแสดงในรูปที่ 2การไหลของอากาศถูกบังคับในส่วนของการทดสอบด้วยวิธีการของ 5.6 กิโลวัตต์พัดลมด้วยอินเวอร์เตอร์ เพื่อหลีกเลี่ยงและลดผลกระทบของการแบ่งปันที่ไม่เหมาะสมหรือไม่เพียงพอการไหลในการทดลอง , อีควอไลเซอร์และเครื่องหนีบผมผสมขึ้นมา ขาเข้าและออก อุณหภูมิทั่วตัวอย่างคอยล์วัดจำนวน 2 T - เทอร์โมคัปเปิ้ลเมนต์ทางเข้าวัดตาข่ายประกอบด้วยสิบสองและในขณะที่เสียบตาข่ายประกอบด้วย 36 และ . เหล่านี้สัญญาณ ข้อมูลนักเรียนรายบุคคลที่บันทึกไว้แล้วเฉลี่ย ในระหว่างการทดสอบอุณหภูมิ ความแปรปรวนของเทอร์โมคัปเปิลเหล่านี้ภายใน 0.2c . นอกจากนี้ , เทอร์โมคัปเปิลทั้งหมดก่อนการสอบเทียบเทอร์โมมิเตอร์โดยควอตซ์กับ 0.01 
c ความละเอียดความดันของขดลวดที่ถูกตรวจพบโดยการทดสอบความแม่นยำแรงดันค่าเซอร์อ่าน 0.1 . สถานีตรวจวัดการไหลของอากาศเป็นรหัสหัวฉีดแบบทดสอบตามมาตรฐาน ASHRAE / [ 28 ] สื่อทำงานในท่อด้านข้างถูกน้ำร้อน ปากน้ำน้ำอุณหภูมิถูกควบคุมโดย thermostat อ่างเก็บน้ำมีความจุได้ถึง 25 กิโลวัตต์ทั้งขาเข้าและขาออกอุณหภูมิวัดจากสอง precalibrated rtds ( อุปกรณ์ ทนอุณหภูมิ pt-100 ) ความถูกต้องของตนเองภายใน 0.05  C อัตราการไหลเชิงปริมาตรน้ำที่ตรวจพบโดยกระแสแม่เหล็กเครื่องวัด 0.002 L / s ความละเอียด
การแปล กรุณารอสักครู่..
 
ภาษาอื่น ๆ
การสนับสนุนเครื่องมือแปลภาษา: กรีก, กันนาดา, กาลิเชียน, คลิงออน, คอร์สิกา, คาซัค, คาตาลัน, คินยารวันดา, คีร์กิซ, คุชราต, จอร์เจีย, จีน, จีนดั้งเดิม, ชวา, ชิเชวา, ซามัว, ซีบัวโน, ซุนดา, ซูลู, ญี่ปุ่น, ดัตช์, ตรวจหาภาษา, ตุรกี, ทมิฬ, ทาจิก, ทาทาร์, นอร์เวย์, บอสเนีย, บัลแกเรีย, บาสก์, ปัญจาป, ฝรั่งเศส, พาชตู, ฟริเชียน, ฟินแลนด์, ฟิลิปปินส์, ภาษาอินโดนีเซี, มองโกเลีย, มัลทีส, มาซีโดเนีย, มาราฐี, มาลากาซี, มาลายาลัม, มาเลย์, ม้ง, ยิดดิช, ยูเครน, รัสเซีย, ละติน, ลักเซมเบิร์ก, ลัตเวีย, ลาว, ลิทัวเนีย, สวาฮิลี, สวีเดน, สิงหล, สินธี, สเปน, สโลวัก, สโลวีเนีย, อังกฤษ, อัมฮาริก, อาร์เซอร์ไบจัน, อาร์เมเนีย, อาหรับ, อิกโบ, อิตาลี, อุยกูร์, อุสเบกิสถาน, อูรดู, ฮังการี, ฮัวซา, ฮาวาย, ฮินดี, ฮีบรู, เกลิกสกอต, เกาหลี, เขมร, เคิร์ด, เช็ก, เซอร์เบียน, เซโซโท, เดนมาร์ก, เตลูกู, เติร์กเมน, เนปาล, เบงกอล, เบลารุส, เปอร์เซีย, เมารี, เมียนมา (พม่า), เยอรมัน, เวลส์, เวียดนาม, เอสเปอแรนโต, เอสโทเนีย, เฮติครีโอล, แอฟริกา, แอลเบเนีย, โคซา, โครเอเชีย, โชนา, โซมาลี, โปรตุเกส, โปแลนด์, โยรูบา, โรมาเนีย, โอเดีย (โอริยา), ไทย, ไอซ์แลนด์, ไอร์แลนด์, การแปลภาษา.

Copyright ©2024 I Love Translation. All reserved.

E-mail: