2.3. Apparatus
A SIA system (MGC Auto-Pret MP-014s, MGC, Japan), which consists of an 8-port selection valve, a 5.0 cm3 syringe pump and a 5 cm3 holding coil, and is computer-controlled using the MGCLMPro ver.2.5 software, was used to deliver the samples and reagents. Square wave ASV (SWASV) measurements were carried out with a potentiostat (eDAQ, Australia) in a thin-layer flow cell (BAS, USA) that consisted of a gasket as a spacer, a D24C8/Nafion SPCE as the working electrode, a silver/silver chloride (Ag/AgCl,3 mol dm_3 KCl) reference electrode, and a stainless-steel tube counter electrode as well as a solution outlet of the flow cell. All SWASV experiments were performed in a laboratory-made Faraday cage at conditioned room temperature. The results from this method in the sample analysis were validated with the ICP-OES.
2.3 อุปกรณ์
ระบบ SIA (MGC อัตโนมัติเสื้อผ้าสำเร็จรูป MP-014s, MGC, ญี่ปุ่น) ซึ่งประกอบด้วยวาล์วเลือก 8 พอร์ต, ปั๊มหลอดฉีดยา 5.0 cm3 และ 5 cm3 ถือม้วนและเป็นคอมพิวเตอร์ควบคุมการใช้ mgclmpro เวอร์ชั่น .2.5 ซอฟต์แวร์ที่ถูกใช้ในการส่งตัวอย่างและสารเคมี ASV คลื่นสี่เหลี่ยม (swasv) วัดได้รับการดำเนินการกับ potentiostat (edaq, ออสเตรเลีย) ในเซลล์ไหลบางชั้น (รูปปั้น,usa) ที่ประกอบไปด้วยปะเก็นเป็น spacer, SPCE d24c8/nafion เป็นขั้วทำงานคลอไรด์สีเงิน / เงิน (ag / ซิลเวอร์คลอไรด์ 3 โมเลกุล dm_3 kcl) ขั้วไฟฟ้าอ้างอิงและสแตนเลสท่อเคาน์เตอร์อิเล็กโทรดที่เป็น ทางออกการแก้ปัญหาของเซลล์ไหล ทุกการทดลอง swasv ได้ดำเนินการในกรงฟาราเดย์ทดลองทำที่อุณหภูมิห้องปรับอากาศผลที่ได้จากวิธีการในการวิเคราะห์ตัวอย่างนี้ได้รับการตรวจสอบกับ ICP OES.