We compared the RSA power derived from electrocardiography (ECG) with PPG at the reference RR derived from capnography. The power of the PPG was significantly higher than that of the ECG (182.42 ± 36.75 dB vs. 162.30 ± 43.66 dB). Further, the mean RR error for PPG was lower than ECG. Both PPG and ECG RR estimation techniques were more powerful and reliable in cases of spontaneous ventilation than when pressure controlled ventilation was used. The analysis of cases containing artifacts in the PPG revealed a significant increase in RR error, a trend that was less pronounced for controlled ventilation. These results indicate that the estimation of RR from the sub-optimally sampled PPG signal is possible and more reliable than from the ECG.