here are several methods used for detecting and measuring particle size or size distribution — light blocking (obscuration), light scattering, Coulter principle and direct imaging.
The light blocking optical particle counter method is typical useful for detecting and sizing particles greater than 1 micrometer in size and is based upon the amount of light a particle blocks when passing through the detection area of the particle counter. This type of technique allows high resolution and reliable measurement.
The light scattering method is capable of detecting smaller sizing particles. This technique is based upon the amount of light that is redirected by a particle passing through the detection area of the particle counter. This redirection is called light scattering. Typical detection sensitivity of the light scattering method is 0.05 micrometre or larger. However, employment of the condensation nuclei counter (CNC) technique would allow a higher detection sensitivity in particle sizes down to nanometre range. A typical application is monitoring of ultrapure water in semiconductor fabrication facilities.
The light blocking method is specified for particle counters that are used for counting in hydraulic and lubricating fluids. Particle counters are used here to measure contamination of hydraulic oil, and therefore allow the user to maintain their hydraulic system, reduce breakdowns, schedule maintenance during no or slow work periods, monitor filter performance, etc. Particle counters used for this purpose typically use ISO Standard 4406:1999 as their reporting standard, and ISO 11171 as the calibration standard. Others also in use are NAS 1638 and its successor SAE AS4059D.
Direct imaging is a technique that uses the light emitted by a laser as a source to illuminate a cell where particle are passing through. The technique does not measures the light blocked by the particles but rather measures the area of the particles functioning like an automated microscope. A pulsed laser diode freezes the particle motion. The light transmitted through the fluid is imaged onto an electronic camera with macro focusing optics. The particles in the sample will block the light and the resulting silhouettes will be imaged onto the digital camera chip.
Other types of particle counters[edit]