Paddles/Electrodes.
Confirm that an adequate supply of disposable defibrillation electrodes (or reusable paddles), ECG electrode, pacing electrodes, or combination defibrillation/pacing electrode (if applicable) are present and that they are within their expiration dates. Confirm that special paddle or electrode (e.g., pediatric, internal) are available, if appropriate. If paddle are used, examine all paddles for physical condition and cleanliness. Alert clinical personnel responsible for the instrument to the presence of dried electrode gel, physiologic fluids, or debris on the paddle surface or handles. Dirty electrodes prevent good electrical contact and can cause burns. Electrode gel or other debris on the insulating portion of the paddle con cause operator shocks. Clean the paddles, if needed, including the electrode surface and handle seams, and make sure that they are completely dry before proceeding with any further testing.
Controls/Switches.
Verify that software setup parameters accessible through hidden or service menus are correctly set for the appropriate application and are consistent for all defibrillators. For units with an automated rhythm recognition program, in addition to any other setup parameters, verify that the correct set of audible and visual user prompts and defibrillation sequences are set and/or that necessary updates have been installed and implemented. Resuscitation guidelines (e.g, American Heart Association) and hospital procedures may vary or change with time. Discuss appropriate settings with the department head and users and ensure that training for any newly introduced protocols is provided.
Examine all controls and switches for physical condition, secure mounting, and correct motion. Where a control should operate against fixed-limit stops, check for proper alignment, as well as positive stopping. During the course of the inspection, check that each control and switch (including membrane switches) performs its proper function.
If the defibrillator has redundant control functions (e.g., a charge button on the oront panel and on a paddle), ensure that both controls function properly. Verify that activating just one paddle discharge button will not cause the defibrillator to discharge. A front-panel discharge button should control only internal paddles (or disposable defibrillator electrodes, on some units) and should not cause discharge when external paddles are connected.
Paddles/Electrodes.
Confirm that an adequate supply of disposable defibrillation electrodes (or reusable paddles), ECG electrode, pacing electrodes, or combination defibrillation/pacing electrode (if applicable) are present and that they are within their expiration dates. Confirm that special paddle or electrode (e.g., pediatric, internal) are available, if appropriate. If paddle are used, examine all paddles for physical condition and cleanliness. Alert clinical personnel responsible for the instrument to the presence of dried electrode gel, physiologic fluids, or debris on the paddle surface or handles. Dirty electrodes prevent good electrical contact and can cause burns. Electrode gel or other debris on the insulating portion of the paddle con cause operator shocks. Clean the paddles, if needed, including the electrode surface and handle seams, and make sure that they are completely dry before proceeding with any further testing.
Controls/Switches.
Verify that software setup parameters accessible through hidden or service menus are correctly set for the appropriate application and are consistent for all defibrillators. For units with an automated rhythm recognition program, in addition to any other setup parameters, verify that the correct set of audible and visual user prompts and defibrillation sequences are set and/or that necessary updates have been installed and implemented. Resuscitation guidelines (e.g, American Heart Association) and hospital procedures may vary or change with time. Discuss appropriate settings with the department head and users and ensure that training for any newly introduced protocols is provided.
Examine all controls and switches for physical condition, secure mounting, and correct motion. Where a control should operate against fixed-limit stops, check for proper alignment, as well as positive stopping. During the course of the inspection, check that each control and switch (including membrane switches) performs its proper function.
If the defibrillator has redundant control functions (e.g., a charge button on the oront panel and on a paddle), ensure that both controls function properly. Verify that activating just one paddle discharge button will not cause the defibrillator to discharge. A front-panel discharge button should control only internal paddles (or disposable defibrillator electrodes, on some units) and should not cause discharge when external paddles are connected.
การแปล กรุณารอสักครู่..

Paddles/Electrodes.
Confirm that an adequate supply of disposable defibrillation electrodes (or reusable paddles), ECG electrode, pacing electrodes, or combination defibrillation/pacing electrode (if applicable) are present and that they are within their expiration dates. Confirm that special paddle or electrode (e.g., pediatric, internal) are available, if appropriate. If paddle are used, examine all paddles for physical condition and cleanliness. Alert clinical personnel responsible for the instrument to the presence of dried electrode gel, physiologic fluids, or debris on the paddle surface or handles. Dirty electrodes prevent good electrical contact and can cause burns. Electrode gel or other debris on the insulating portion of the paddle con cause operator shocks. Clean the paddles, if needed, including the electrode surface and handle seams, and make sure that they are completely dry before proceeding with any further testing.
Controls/Switches.
Verify that software setup parameters accessible through hidden or service menus are correctly set for the appropriate application and are consistent for all defibrillators.หน่วยจังหวะกับอัตโนมัติโปรแกรมรู้จำนอกจากพารามิเตอร์ใด ๆการติดตั้งอื่น ๆ ตรวจสอบว่า การตั้งค่าเสียงและภาพให้ถูกต้องของผู้ใช้และการกระตุกหัวใจด้วยไฟฟ้าดับเป็นชุดและ / หรือการปรับปรุงที่จำเป็นได้รับการติดตั้งและใช้ แนวทางการกู้ชีพ ( เช่น สมาคมหัวใจอเมริกัน ) และขั้นตอนของโรงพยาบาลอาจเปลี่ยนแปลงหรือเปลี่ยนแปลงกับเวลาหารือเกี่ยวกับการตั้งค่าที่เหมาะสมกับหัวหน้าภาควิชา และผู้ใช้ และมั่นใจว่า การเข้าชมใด ๆโปรโตคอลที่ให้ไว้ และสวิทช์สำหรับการควบคุม
ตรวจสภาพทางกายภาพ , การติดตั้ง และแก้ไข การเคลื่อนไหว ที่ควบคุมงานกับขีด จำกัด ของการแก้ไข ควรหยุดการตรวจสอบที่เหมาะสม รวมทั้ง บวกหยุด ในระหว่างการตรวจสอบตรวจสอบและควบคุมแต่ละสวิตช์ ( รวมถึงสวิทช์เมมเบรน ) ทำหน้าที่ที่เหมาะสม ถ้าเครื่องมีฟังก์ชั่นการควบคุม
ซ้อน ( เช่น ปุ่มชาร์จบนแผง oront และพาย ) , ตรวจสอบให้แน่ใจว่าทั้งการควบคุมฟังก์ชันอย่างถูกต้อง ตรวจสอบว่าเปิดใช้งานปุ่มปลดแค่ใบพัดจะไม่ทำให้เครื่องที่จะปล่อยแผงปุ่มควบคุมด้านหน้าจะปล่อยเพียงพายภายใน ( หรือขั้วไฟฟ้าเครื่องผ้าอ้อมในบางหน่วย ) และไม่ควรให้จำหน่ายเมื่อ paddles ภายนอกเชื่อมต่อ .
การแปล กรุณารอสักครู่..
