Report HALM MAXWELL tester light intensity calibration work instructions.
Operating procedures Calibration.
1.ISC inspection is not standardized.
1.1 Check Flasher XF2 (1000±5)
1.2 Checking the straight place an of the bus bar.
1.3 Checking the speed of the conveyor belt
1.4Checking the speed of the cycle time.
" 1.5Checking the calibration standard (Check cell area refernece list Mono 156.75*156.75=244.32 cm²
Poly156*156=243.33 cm² ISC;±0.01A Voc;±0.002V FF;(%)±2% EFF;(%)±2% Temperature (25±2 Cº))"
1.6 Checking The sensors Cleaning
2.VOC inspection is not standardized.
2.1 Check Temperature (25±2 Cº)
3. FF inspection is not standardized.
3.1Checking the scope needle
3.2Checking the scope of putting needle on cells.
3.3 Check the the power cord abnormalities.