Cyclone Calibration Adapter
Cat. No. 225-01-03
The calibration adapter is for monitoring the flow through a fi lter/cyclone. It is designed to fi t cyclones Cat. No. 225-01-01 and
225-01-02.
Assemble the
fi lter/cyclone assembly in the normal way and connect
to the sample pump.
Moisten the O-ring inside the large end of the adapter and push
onto the enlarged section of the cyclone extension. Connect the
small end to a
fl owmeter or other calibration instrument such as
the Defender 717 Series. Use 1/4 in ID tubing to make the connection
and maintain as short as possible. (See illustration on reverse side.
)
Turn on pump and read
fl ow from the calibrator.