check furnace checkboxes ( all three must be checked)
install upgrade kit17 for gas outlets and check integrity of moisture absorbent.
reduce helium intel pressure to 1.0 to 1,5 bar to save helium carrier gas.
take out ash insert overnight and close connectors again gastight.
make sure preventive degassing is switched on
keep system in work mode and uncheck furnace checkboxes
for combustion reduction and degassing
take out ash insert
wait for cooling down untill combustion furnace temperature reaches 500
dumatherm must be cooled down to minimum 500 or lowervwith gases on before the instument is switched off completely
after reching the 500 all gas outlets must be closed with the supplied brass nuts as shown in the picture
alternatively and better upgrade should be installed to both outlets degas and output. this prevents water coming into the system.