(example)
Notice: The calibration of the instrument must be prepared carefully. Pay attention to a homogeneous sample material without dirt and green grain portions. Do not change the calibration after one measurement only. Carry out at least three measurements with the HE lite and calculate the average. Calculate the average deviation to the comparative instrument.
The effects of increasing differences with increasing moisture contents can be corrected by means of the multi placative factor GAIN. For this purpose at least two samples, if possible, of different moisture contents are required, since otherwise the effects cannot be determined exactly.
Since the procedure is complicated(determination of the gain of a straight line) we recommend to consult the manufacturer before changing the GAIN!
Change the factor by means of the arrow keys and terminate the input by pressing: