Because the respiratory system provides the aeromechanical
drive to the larynx and upper-airway structures
for speech production, SCI can impair speech. Speech is
often affected by injuries to the cervical cord (and is typically
normal with lower injuries). When cervical injury
spares diaphragm function, common speech signs include
low loudness (due to impaired alveolar pressure-generation
capability) and short phrases (due to small VT).78 In
such cases, an abdominal binder may improve speech.79
When cervical injury impairs diaphragm function, a ventilator
is required and speech is substantially altered. The
specific nature of ventilator-assisted speech depends on
the type of ventilator used. Discussion here will focus on
invasive positive-pressure ventilation and ventilation with
a phrenic-nerve pacer. More detailed accounts of these
forms of ventilator-assisted speech are available elsewhere.
80–82
Speech produced with invasive positive-pressure ventilation
delivered via tracheostomy is substantially different
from normal speech, primarily because the ventilator-delivered
air enters below the larynx. If the tracheostomy
tube cuff is inflated (and the tube is unfenestrated), the
patient will not be able to speak at all. Therefore, the first
and most critical step toward allowing a patient to speak is
to deflate the cuff. This can be done safely by increasing
the ventilator-delivered VT to compensate for air loss
through the larynx. A typical error is to choose a VT that
is too small to maintain a constant PaCO2
. Commonly, we
have used VT as high as 1.2 L. This is not as large as first
appears. The inspiratory capacity of an adult of average
size is about 3 L, so a 1.2-L breath does not bring the
subject even close to total lung capacity. In addition, the
deflated cuff allows gas to leak out of the mouth for speech
production, so the effective VT is much less than 1.2 L.
Deflation of the cuff makes it possible for the patient to
communicate relatively easily with family, friends, and
hospital staff. Nevertheless, tracheostomy-ventilation
speech is usually abnormal and is characterized by short
phrases, long pauses, poor voice quality, and variable loudness.
83 These abnormal speech characteristics are closely
linked to the abnormal tracheal pressure associated with
tracheostomy ventilation. As shown in Figure 2, tracheostomy-
ventilation tracheal pressure (under a volume-controlled
condition) rises quickly to a high peak, falls quickly,
and remains below the minimum pressure required to vibrate
the vocal folds until the next inspiration. This is in
stark contrast to the steady and low tracheal pressure associated
with normal speech production.
Fortunately, tracheal pressure can be modified with ventilator
adjustments to improve speech. Perhaps the most
simple and successful set of ventilator adjustments is the
combination of prolonged inspiratory time and positive
end-expiratory pressure (PEEP).80,84 Prolonged inspiratory
time (or decreased inspiratory flow) increases the time that
speech can be produced during inspiration, and PEEP increases
the time that speech can be produced during ex-
Because the respiratory system provides the aeromechanical
drive to the larynx and upper-airway structures
for speech production, SCI can impair speech. Speech is
often affected by injuries to the cervical cord (and is typically
normal with lower injuries). When cervical injury
spares diaphragm function, common speech signs include
low loudness (due to impaired alveolar pressure-generation
capability) and short phrases (due to small VT).78 In
such cases, an abdominal binder may improve speech.79
When cervical injury impairs diaphragm function, a ventilator
is required and speech is substantially altered. The
specific nature of ventilator-assisted speech depends on
the type of ventilator used. Discussion here will focus on
invasive positive-pressure ventilation and ventilation with
a phrenic-nerve pacer. More detailed accounts of these
forms of ventilator-assisted speech are available elsewhere.
80–82
Speech produced with invasive positive-pressure ventilation
delivered via tracheostomy is substantially different
from normal speech, primarily because the ventilator-delivered
air enters below the larynx. If the tracheostomy
tube cuff is inflated (and the tube is unfenestrated), the
patient will not be able to speak at all. Therefore, the first
and most critical step toward allowing a patient to speak is
to deflate the cuff. This can be done safely by increasing
the ventilator-delivered VT to compensate for air loss
through the larynx. A typical error is to choose a VT that
is too small to maintain a constant PaCO2
. Commonly, we
have used VT as high as 1.2 L. This is not as large as first
appears. The inspiratory capacity of an adult of average
size is about 3 L, so a 1.2-L breath does not bring the
subject even close to total lung capacity. In addition, the
deflated cuff allows gas to leak out of the mouth for speech
production, so the effective VT is much less than 1.2 L.
Deflation of the cuff makes it possible for the patient to
communicate relatively easily with family, friends, and
hospital staff. Nevertheless, tracheostomy-ventilation
speech is usually abnormal and is characterized by short
phrases, long pauses, poor voice quality, and variable loudness.
83 These abnormal speech characteristics are closely
linked to the abnormal tracheal pressure associated with
tracheostomy ventilation. As shown in Figure 2, tracheostomy-
ventilation tracheal pressure (under a volume-controlled
condition) rises quickly to a high peak, falls quickly,
and remains below the minimum pressure required to vibrate
the vocal folds until the next inspiration. This is in
stark contrast to the steady and low tracheal pressure associated
with normal speech production.
Fortunately, tracheal pressure can be modified with ventilator
adjustments to improve speech. Perhaps the most
simple and successful set of ventilator adjustments is the
combination of prolonged inspiratory time and positive
end-expiratory pressure (PEEP).80,84 Prolonged inspiratory
time (or decreased inspiratory flow) increases the time that
speech can be produced during inspiration, and PEEP increases
the time that speech can be produced during ex-
การแปล กรุณารอสักครู่..
เพราะระบบการหายใจให้ aeromechanical
ขับกล่องเสียงและบนโครงสร้างการบิน
การผลิตการพูด , วิทย์สามารถบั่นทอนประสิทธิภาพการพูด การพูด คือ มักจะได้รับผลกระทบจาก
การบาดเจ็บที่ไขสันหลัง ( และมักจะ
ปกติกับการบาดเจ็บลดลง ) เมื่อปากมดลูกบาดเจ็บ
อะไหล่ฟังก์ชันไดอะแฟรม , ป้ายคำพูดทั่วไปรวมถึงความดังต่ำ ( เนื่องจาก
รุ่นแจ่มใสบกพร่องความสามารถ ) และประโยคสั้น ๆเนื่องจาก VT ) 78 ใน
กรณีดังกล่าว เป็นเครื่องผูกท้องอาจปรับปรุงการพูด . 79
เมื่อปากมดลูกการบาดเจ็บทำลายการทำงานไดอะแฟรม , เครื่องช่วยหายใจ
ถูก ต้อง และ การพูด มีการเปลี่ยนแปลงอย่างมาก
ลักษณะจำเพาะของเครื่องช่วยหายใจช่วยการพูดขึ้นอยู่กับ
ชนิดของเครื่องช่วยหายใจที่ใช้ การสนทนานี้จะเน้นที่การระบาย และระบายแรงดันบวก
ด้วยเส้นประสาทกะบังลมเป็นเครื่ . รายละเอียดเพิ่มเติมบัญชีเหล่านี้
รูปแบบของเครื่องช่วยหายใจช่วยการพูดมีอยู่ที่อื่น ๆ .
คำพูด 82 80 ) ผลิตด้วยแรงดันบวกผ่านทางวิธีการระบาย
ส่งอสัญแตกต่างอย่างมาก
จากการพูดธรรมดา เป็นหลัก เพราะเครื่องช่วยหายใจส่ง
อากาศเข้าสู่ใต้กล่องเสียง ถ้าอสัญ
ข้อมือพอง ( หลอดและท่อเป็น unfenestrated ) ,
ผู้ป่วยจะไม่สามารถพูดได้ทั้งหมด ดังนั้น แรก และสำคัญที่สุดขั้นตอนต่อ
ให้คนไข้พูดจาที่ทำให้ข้อมือ . นี้สามารถทำได้โดยการใช้อย่างปลอดภัย
ส่ง VT เพื่อชดเชยการสูญเสีย
ลมผ่านกล่องเสียง ข้อผิดพลาดโดยทั่วไปจะเลือก VT ที่
มีขนาดเล็กเกินไปที่จะรักษา
paco2 คงที่โดยทั่วไปเรา
ใช้ VT สูงถึง 1.2 ลิตร ไม่ใช่ขนาดใหญ่เป็นครั้งแรก
ปรากฏ ความสามารถในการของผู้ใหญ่ ขนาดเฉลี่ย
ประมาณ 3 ลิตร ดังนั้น 1.2-l ลมหายใจไม่ได้เอา
เรื่องแม้แต่ทั้งหมดความจุปอด นอกจากนี้ กิ่วข้อมือช่วยให้แก๊สรั่ว
พูดออกจากปากเพื่อการผลิต ดังนั้น VT มีประสิทธิภาพมากน้อยกว่า 1.2 L .
ภาวะเงินฝืดของข้อมือได้ทำให้มันเป็นไปได้สำหรับผู้ป่วย
ติดต่อค่อนข้างง่ายกับครอบครัว , เพื่อน , และ
พนักงานโรงพยาบาล อย่างไรก็ตาม การเจาะคอระบาย
การพูดมักจะผิดปกติและมีลักษณะสั้น
วลี หยุดยาว คุณภาพเสียงไม่ดี และตัวแปรดัง .
83 การพูดลักษณะที่ผิดปกติเหล่านี้อย่างใกล้ชิด
เชื่อมโยงไปยังความผิดปกติของหลอดลมความดันที่เกี่ยวข้องกับ
อสัญระบายอากาศ ดังแสดงในรูปที่ 2 , การเจาะคอระบายความดัน ( -
ว่าภายใต้สภาวะควบคุมปริมาตร
) เพิ่มขึ้นอย่างรวดเร็วถึงจุดสูงสุดสูงตกลงมาอย่างรวดเร็ว
และยังคงอยู่ด้านล่างแรงดันต่ำต้องสั่น
เท่าเสียงจนเกิดแรงบันดาลใจต่อไป นี่คือ
ตรงกันข้ามกับ steady และความดันต่ำ สตาร์คว่า
ที่เกี่ยวข้องกับการผลิตการพูดปกติ
โชคดี สามารถแก้ไขได้ด้วยการปรับแรงดันที่หลอดลมเครื่องช่วยหายใจ
ปรับปรุงการพูด บางทีที่สุด
ง่ายและประสบความสำเร็จชุดของการปรับใช้คือการรวมกันของการเวลานาน
จบและบวก Expiratory ความดัน ( แอบดู ) 80,84 inspiratory
นานเวลา ( หรือลดลงการไหล ) เพิ่มเวลาที่
การพูดสามารถผลิตระหว่างแรงบันดาลใจและเพิ่มขึ้นเมียงมอง
เวลาปราศรัยนั้นสามารถผลิตได้ในช่วงอดีต
การแปล กรุณารอสักครู่..