2. Minor load application
Turn the hand-wheel to apply the minor load, when the indicating bar reaches the left edge of the set
area, the minor load is successfully applied. The length gauge reading should be approximately 600μm.
3. Major load selection
Turn the major load selection dial and observe the screen to check if the major load is correctly
displayed according to the selection on the dial.
4. Major load application
Press the OK key to apply the major load after you apply the minor load. The motor drives the power
lever downwards and upwards. If the major load is being applied and removed correctly, you should
observe the length gauge reading decrease first and then increase.