Carefully remove padded shells or sealing ring from the haematocrit rotor.
2. Soak haematocrit rotor, lid and padded shells or sealing ring in cold water, until the blood residues are
completely dissolved.
3. Perform a suitable disinfection. The application of disinfection substances on the lid is to be avoided as the
inscription can dissolve.
Ingredients of suitable disinfectants: glutaraldehyde, propanol, ethyl hexanol, anionic tensides, corrosion
inhibitors.
Ingredients of suitable agents for removing radioactive contaminants: anionic tensides, non-ionic tensides,
polyhydrated ethanol.
4. Rinse the components in cold water and dry them.
5. They can now be replaced into the centrifuge.