It should be stressed that the Q-meter measures the effective inductance and the Q
is determined as the ratio of the effective inductive reactance to the effective series real
part of the loop impedance. At HF and VHF, the effective parallel capacitance of the coil
acts to reduce the effective Q (equation 5.3).