• Switch button CCW/CW Switch 9 to the direction of rotation, which the manual
applicator is connected to the tube.
Direction of rotation: CW – clockwise CCW – counter clockwise
• With the Speed Adjustment Knob 3 choose a speed.
• Initiate a start to check the dispensed quantity.
If the dispensed quantity is too small:
• Increase the dispensing time and/or the speed of the drive. Check the dispensed
quantity again. Repeat this sequence until the desired dispensed quantity is
achieved.
If the dispensed quantity is too high:
• Decrease the dispensing time and/or the speed of the drive. Check the dispensed
quantity again. Repeat this sequence until the desired dispensed quantity is
achieved.
If the desired dispensed quantity is not within reach:
• Change the feedline to a smaller/bigger size and repeat the above steps.
R.C. MODE:
• Switch button Operation Mode Switch 2 to R.C.
The dispensing takes place as long as:
– The foot switch is pressed,
– The finger switch is pressed or
– A start signal from a higher-ranking controller is present.