Steam sterilisation using autoclaves or retorts is a widely used thermal process in the food processing and
pharmaceutical industries. Unfortunately, sterilisation, using steam, consumes a significant amount of
energy and is known to cause unwanted peak energy demand when a number of autoclaves operate in
parallel. The objective of this paper is to extend the previously published numerical methodology developed
to simulate in detail the thermal energy consumption of an industrial sized empty autoclave used
for steam sterilisation, to include products, in this case, intravenous solution packed in plastic pouches
and a steam flow controller. The external parameters supplied to the numerically modelled controller
are the maximum steam flow rate and the autoclave temperature as a function of time which it needs
to maintain. The numerical model is then used to predict the actual transient temperature and pressure
profiles and the details of the mass transfers in the autoclave during a sterilisation cycle, the transient
temperature distribution within the products as well as details of the transient thermal energy consumption.
The results from numerical modelling were validated with measurements obtained under actual
operating conditions. The simulated total steam consumption was within 3% of the measured data. A
reduction of 8% in steam consumption was obtained due to insulation on the outer walls.