a quaternary pump with integrated vacuum degasser (G1311C),
autosampler (G1329B), column oven (G1316A) and fluorescence
detector (G1321A). The separation, identification and quantification
steps of OTA were carried out according to the EN 14133:2009
method. Briefly, the separation was performed on a Zorbax column
SB-C18, 4.6 × 150 mm, 5-Micron (Agilent, Santa Clara, CA, USA),
under isocratic conditions. The mobile phase was a mixture of acetoni-
trile, water and acetic acid (49.5:49.5:1 v/v/v, respectively). The flow
was set to 1 mL/min and the column oven temperature was set to
25 °C. The excitation and the emission wavelengths were set to 333
and 460 nm, respectively. OTA was identified by comparing the reten-
tion time of the relevant peak on samples with the one from the stan-
dard (5.9 min). The quantification of OTA was performed using a
linear external calibration in a range from 0.6 to 60 ng/mL, with a
mean correlation coefficient of 0.99995.
ปั๊ม quaternary กับ degasser เครื่องดูดฝุ่นรวม (G1311C),autosampler (G1329B), คอลัมน์เตา (G1316A) และ fluorescenceเครื่องตรวจจับ (G1321A) แยก identification และ quantificationขั้นตอนการ OTA ได้ดำเนินการตาม 14133:2009 น้ำวิธีการ ทำ Briefly แยกตามคอลัมน์ ZorbaxSB-C18, 4.6 × 150 mm, 5 ไมครอน (Agilent ซานตาคลารา CA, USA),ภายใต้เงื่อนไข isocratic คิด เฟสเคลื่อนที่มีส่วนผสมของ acetoni-trile น้ำและกรดอะซิติก (49.5:49.5:1 v/v/v ตามลำดับ) flowถูกตั้งค่าเป็น 1 mL/min และอุณหภูมิเตาอบคอลัมน์ถูกตั้งค่าให้25 องศาเซลเซียส การในการกระตุ้นและความยาวคลื่นที่ปล่อยก๊าซถูกตั้งค่า 333และ 460 nm ตามลำดับ โอตะถูก identified โดยเปรียบเทียบ reten-เวลาสูงสุดที่เกี่ยวข้องกับตัวอย่างด้วยจากสแตน - สเตรชันdard (5.9 นาที) Quantification ของโอตะถูกดำเนินการโดยใช้การปรับเทียบภายนอกเชิงเส้นในช่วงตั้งแต่ 0.6 ถึง 60 ng/mL มีการหมายถึง ความสัมพันธ์ของ coefficient ของ 0.99995