5.2.6 Completing a calibration
The T100 & TC120 allow a two point calibration to be completed. The calibration menu
can be accessed by simultaneously pressing the F and S buttons for about 5 seconds.
The calibration temperatures are constrained by the temperature limits of the liquid type
setting. Calibration should be carried out using a traceable reference thermometer with
an accuracy of at least 0.1°C. This thermometer should be held securely in the centre of
the bath or vessel.