Battery/Charger. For difibrillators with removable batteries that are  การแปล - Battery/Charger. For difibrillators with removable batteries that are  ไทย วิธีการพูด

Battery/Charger. For difibrillators

Battery/Charger.
For difibrillators with removable batteries that are charged in a separate charger, inspect the physical condition of the batteries and battery connectors, and verify that the batteries in the charger are properly installed and that the charging or ready light is on.
Verify that the line cord is firmly attached at the wall outlet and instrument end. For defibrillators that can be connected to AC line power for the purposes of charging, verify that the unit is plugged into alive AC outlet, the line cord is firmly attached at the wall outlet and instrument end, the charger is attached to the defibrillator (i.e., the charger cable is attached or the defibrillator is firmly seated into the charging stand or mount) and that the Indicators/Displays.
During the course of the inspection, confirm the operation of all lights, indicators, meters, and visual displays on the monitor and the charger (if appropriate). For units with an ECG monitor, observe a simulated ECG signal on the display, and verify compliance with the following criteria: The baseline should be flat and without 50/60 Hz interference or other noise with the ECG simulator attached.
All portions of a simulated ECG waveform should be clear and visible, including the P-wave and QRS.
Ambient light should not affect the visibility of the trace. ( If monitor reflects ambient light, making the ECG difficult to see, try adjusting the display brightness and/or contrast.)
For units with pacing capabilities, connect the monitoring leads to an E4CG simulator or pacemaker analyzer. Connect the pacing cable to an appropriate test load or a pacemaker analyzer, start the pacemaker, and print a strip. Verify that the pacing maker rate varies as the pacing rate control is adjusted.
0/5000
จาก: -
เป็น: -
ผลลัพธ์ (ไทย) 1: [สำเนา]
คัดลอก!
Battery/Charger. For difibrillators with removable batteries that are charged in a separate charger, inspect the physical condition of the batteries and battery connectors, and verify that the batteries in the charger are properly installed and that the charging or ready light is on.Verify that the line cord is firmly attached at the wall outlet and instrument end. For defibrillators that can be connected to AC line power for the purposes of charging, verify that the unit is plugged into alive AC outlet, the line cord is firmly attached at the wall outlet and instrument end, the charger is attached to the defibrillator (i.e., the charger cable is attached or the defibrillator is firmly seated into the charging stand or mount) and that the Indicators/Displays. During the course of the inspection, confirm the operation of all lights, indicators, meters, and visual displays on the monitor and the charger (if appropriate). For units with an ECG monitor, observe a simulated ECG signal on the display, and verify compliance with the following criteria: The baseline should be flat and without 50/60 Hz interference or other noise with the ECG simulator attached.All portions of a simulated ECG waveform should be clear and visible, including the P-wave and QRS.Ambient light should not affect the visibility of the trace. ( If monitor reflects ambient light, making the ECG difficult to see, try adjusting the display brightness and/or contrast.) For units with pacing capabilities, connect the monitoring leads to an E4CG simulator or pacemaker analyzer. Connect the pacing cable to an appropriate test load or a pacemaker analyzer, start the pacemaker, and print a strip. Verify that the pacing maker rate varies as the pacing rate control is adjusted.
การแปล กรุณารอสักครู่..
ผลลัพธ์ (ไทย) 2:[สำเนา]
คัดลอก!
Battery/Charger.
For difibrillators with removable batteries that are charged in a separate charger, inspect the physical condition of the batteries and battery connectors, and verify that the batteries in the charger are properly installed and that the charging or ready light is on.
Verify that the line cord is firmly attached at the wall outlet and instrument end. For defibrillators that can be connected to AC line power for the purposes of charging, verify that the unit is plugged into alive AC outlet, the line cord is firmly attached at the wall outlet and instrument end, the charger is attached to the defibrillator (i.e., the charger cable is attached or the defibrillator is firmly seated into the charging stand or mount) and that the Indicators/Displays.
During the course of the inspection, confirm the operation of all lights, indicators, meters, and visual displays on the monitor and the charger (if appropriate). For units with an ECG monitor, observe a simulated ECG signal on the display, and verify compliance with the following criteria: The baseline should be flat and without 50/60 Hz interference or other noise with the ECG simulator attached.
All portions of a simulated ECG waveform should be clear and visible, including the P-wave and QRS.
Ambient light should not affect the visibility of the trace. ( If monitor reflects ambient light, making the ECG difficult to see, try adjusting the display brightness and/or contrast.)
For units with pacing capabilities, connect the monitoring leads to an E4CG simulator or pacemaker analyzer. Connect the pacing cable to an appropriate test load or a pacemaker analyzer, start the pacemaker, and print a strip. Verify that the pacing maker rate varies as the pacing rate control is adjusted.
การแปล กรุณารอสักครู่..
ผลลัพธ์ (ไทย) 3:[สำเนา]
คัดลอก!
Battery/Charger.
For difibrillators with removable batteries that are charged in a separate charger, inspect the physical condition of the batteries and battery connectors, and verify that the batteries in the charger are properly installed and that the charging or ready light is on.
Verify that the line cord is firmly attached at the wall outlet and instrument end. For defibrillators that can be connected to AC line power for the purposes of charging, verify that the unit is plugged into alive AC outlet, the line cord is firmly attached at the wall outlet and instrument end, the charger is attached to the defibrillator (i.e., the charger cable is attached or the defibrillator is firmly seated into the charging stand or mount) and that the Indicators/Displays.
During the course of the inspection, confirm the operation of all lights, indicators, meters, and visual displays on the monitor and the charger (if appropriate). For units with an ECG monitor, observe a simulated ECG signal on the display,และตรวจสอบการปฏิบัติตามเกณฑ์ต่อไปนี้ : พื้นฐานควรจะแบนและไม่มี 50 / 60 Hz รบกวนหรือเสียงอื่น ๆที่มี ECG Simulator แนบ .
ทุกส่วนของระบบจำลองสัญญาณควรจะชัดเจน และมองเห็นได้ รวมถึง p-wave และ QRS .
แสงไม่ควรส่งผลกระทบต่อทัศนวิสัยในการติดตาม ( หากจอภาพสะท้อนแสง ทำให้ระบบยากที่จะเห็น try adjusting the display brightness and/or contrast.)
For units with pacing capabilities, connect the monitoring leads to an E4CG simulator or pacemaker analyzer. Connect the pacing cable to an appropriate test load or a pacemaker analyzer, start the pacemaker, and print a strip. Verify that the pacing maker rate varies as the pacing rate control is adjusted.
การแปล กรุณารอสักครู่..
 
ภาษาอื่น ๆ
การสนับสนุนเครื่องมือแปลภาษา: กรีก, กันนาดา, กาลิเชียน, คลิงออน, คอร์สิกา, คาซัค, คาตาลัน, คินยารวันดา, คีร์กิซ, คุชราต, จอร์เจีย, จีน, จีนดั้งเดิม, ชวา, ชิเชวา, ซามัว, ซีบัวโน, ซุนดา, ซูลู, ญี่ปุ่น, ดัตช์, ตรวจหาภาษา, ตุรกี, ทมิฬ, ทาจิก, ทาทาร์, นอร์เวย์, บอสเนีย, บัลแกเรีย, บาสก์, ปัญจาป, ฝรั่งเศส, พาชตู, ฟริเชียน, ฟินแลนด์, ฟิลิปปินส์, ภาษาอินโดนีเซี, มองโกเลีย, มัลทีส, มาซีโดเนีย, มาราฐี, มาลากาซี, มาลายาลัม, มาเลย์, ม้ง, ยิดดิช, ยูเครน, รัสเซีย, ละติน, ลักเซมเบิร์ก, ลัตเวีย, ลาว, ลิทัวเนีย, สวาฮิลี, สวีเดน, สิงหล, สินธี, สเปน, สโลวัก, สโลวีเนีย, อังกฤษ, อัมฮาริก, อาร์เซอร์ไบจัน, อาร์เมเนีย, อาหรับ, อิกโบ, อิตาลี, อุยกูร์, อุสเบกิสถาน, อูรดู, ฮังการี, ฮัวซา, ฮาวาย, ฮินดี, ฮีบรู, เกลิกสกอต, เกาหลี, เขมร, เคิร์ด, เช็ก, เซอร์เบียน, เซโซโท, เดนมาร์ก, เตลูกู, เติร์กเมน, เนปาล, เบงกอล, เบลารุส, เปอร์เซีย, เมารี, เมียนมา (พม่า), เยอรมัน, เวลส์, เวียดนาม, เอสเปอแรนโต, เอสโทเนีย, เฮติครีโอล, แอฟริกา, แอลเบเนีย, โคซา, โครเอเชีย, โชนา, โซมาลี, โปรตุเกส, โปแลนด์, โยรูบา, โรมาเนีย, โอเดีย (โอริยา), ไทย, ไอซ์แลนด์, ไอร์แลนด์, การแปลภาษา.