The volatile compounds were isolated, identified and quantified on a Shimadzu GC-17A gas chromatograph (Shimadzu Corporation, Kyoto, Japan), coupled to a Shimadzu mass spectrometer detector (GC–MS QP-5050A). The GC–MS system was equipped with the same column used in the GC analysis and with the same temperature programme. Analyses were carried out using helium as carrier gas at a flow rate of 1 mL/min, at a split ratio of 1:10. Diluted samples (1/10 cyclopentane, v/v) of 0.3 μL of the extracts were injected. Mass spectra were obtained by electron ionization (EI) at 70 eV, using a spectral range of 45–450 m/z. Most of the compounds were identified using two different analytical methods: (a) KI, Kovats Indexes in reference to n-alkanes (C8–C32) (NIST, 2010); and (b) mass spectra (authentic chemicals and Wiley spectral library collection). Identification was considered tentative when based on mass spectral data alone.
สารประกอบที่ระเหยได้แยก , ระบุและ quantified ในแก๊สโครมาโตกราฟ gc-17a Shimadzu ( Shimadzu Corporation , เกียวโต , ญี่ปุ่น ) คู่กับ Shimadzu แมสสเปกโทรมิเตอร์ตรวจจับ ( GC ) นางสาว qp-5050a ) GC - MS ระบบติดตั้งแบบคอลัมน์วิเคราะห์ GC กับโปรแกรมที่อุณหภูมิเดียวกัน วิเคราะห์ข้อมูล โดยใช้ฮีเลียมเป็นแก๊สตัวพา ที่อัตราการไหล 2 มล. / นาที , ที่แบ่งอัตราส่วน 1 : 10 . เจือจางตัวอย่าง ( 1 / 10 ไซโคลเพนเทน , V / V ) 0.3 μลิตรสารสกัดถูกฉีดเข้าไป มวลสเปกตรัมที่ได้จากอิเล็กตรอนอิสระ ( EI ) ที่ 70 ปีที่ผ่านมา โดยใช้สเปกตรัมช่วง 45 – 450 m / Z . ที่สุดของสารประกอบที่ถูกระบุว่าใช้สองวิธีที่แตกต่างกัน : ( ) กิ kovats ดัชนีอ้างอิง n-alkanes ( ดอง ) c32 ( NIST ) , 2010 ) ; และ ( B ) มวล สเปกตรัม ( สารเคมีจริงและนิ่งสเปกตรัมห้องสมุดคอลเลกชัน ) การพิจารณากำหนดเวลาขึ้นอยู่กับมวลสเปกตรัมข้อมูลเพียงอย่างเดียว