PROCEDURE FOR HI 2314
• Pour a small quantity of the appropriate
calibration solution, e.g. HI 7030
(12.88 mS/cm @ 25°C), into a plastic
beaker.
• Immerse the probe in the solution while
completely submerging the holes of the
sleeve. Tap the probe lightly on the bottom
of the beaker to remove any air bubbles
trapped inside the sleeve. Measure the
solution temperature with a Checktemp
or another accurate thermometer.
• Press and hold the EC/°C key to select
temperature setting.
• Adjust the Temperature knob to display the temperature of the
solution on the LCD.