Troubleshooting
1) Verify the test equipment is operating properly, check or service the battery (if battery operated); following the manufacturer’s operating instructions for proper operation.
2) Check that the test equipment is properly grounded.
3) Verify there are no large conductive objects in the area that could cause ion attraction to their surface.
4) Check that items in the work area are not blocking the ionizer airflow.
5) Check ionizer fans for proper operation.
6) Check air filters for clogging with dust and dirt. PM team clean or replace air filters as necessary.
7) Check emitter points (they should be clean and not bent or broken).
8) Check electrical equipment grounds for grounding integrity.
9) If applicable, PM team adjust the offset voltage of the ionizer following the manufacturer’s instructions.
10) If require, verify Charge Plate Monitor with specific reference voltage (Ex. 0V. 5V. 10V. and -5V. -10V.) at Calibration Lab.