The precision and LSC values provided by the manufacturer cannot be applied to clinical bone densitometry centers because of the differences in the patients being tested and the technologist performing the test. Because harmful errors in clinical management may occur from incorrectly interpreting serial BMD tests, it is recommended that every DXA technologist conduct a precision assessment and calculate the LSC for each measurement site and DXA instrument used. Precision assessment provides direct benefit to patients by allowing clinicians to make clinical decisions based on genuine change or stability of BMD. The patient-care benefits of precision assessment outweigh the risk of exposure to trivial doses of ionizing radiation.